Series TT01 Digital Cap Torque Testers
User’s Guide
19
8. After displaying “PLEASE WAIT…” the display will appear as follows:
Remove the torque applied in Step 7, leave the fixtures in place, then press
ZERO
.
9. The display will appear as follows:
Use the
UP
and
DOWN
keys to adjust the torque value as required. The torque values default to
even increments, as indicated by the previously entered number of data points described in Step
1. Then press
ENTER
.
Repeat the above step for the number of data points selected.
10. After all the clockwise calibration points have been completed, the display will appear as follows:
Press
ENTER
.
11. At the completion of the counter-clockwise calibration, the display will appear as follows:
To save the calibration information, select “SAVE & EXIT”. To exit without saving the data select
“EXIT W/O SAVING”.
12. Any errors are reported by the following screens:
CALIBRATION
CLOCKWISE
ENSURE NO LOAD
THEN PRESS ZERO
CALIBRATION
COMPLETE
SAVE & EXIT
EXIT W/O SAVING
CALIBRATION
CLOCKWISE COMPLETE
REVERSE DIRECTION
FOR CCW
Attach necessary
weight fixtures.
THEN PRESS ENTER
CALIBRATION
CLOCKWISE
APPLY LOAD
1 OF 5
ENTER LOAD:
10.000 LBFIN
THEN PRESS ENTER