Series TT02 Digital Torque Tool Testers
User’s Guide
24
To re-enter the password, press ESC to exit to the home screen. Then, access the desired function and
enter the password again when prompted.
If the password has been misplaced, it can be reset. Press
ENTER
to generate a
request code
. The
request code
must be supplied to Mark-10 or a distributor, who will then provide a corresponding
activation code
. Enter the
activation code
to disable the password.
15 OTHER SETTINGS
15.1 Automatic Shutoff
The tester may be configured to automatically power off following a period of inactivity while on battery
power. Inactivity is defined as the absence of any key presses or load changes of 100 counts or less. To
access these settings, select
Automatic Shutoff
from the menu. The display appears as follows:
Select
Disabled
to disable automatic shutoff. Select
Enabled
to enable it. The length of time of inactivity
is programmed in minutes via the
Set Minutes
parameter. Available settings:
5-30,
in 5 minute
increments.
Note:
If the AC adapter is plugged in, the tester will ignore the
Automatic Shutoff
setting and remain
powered on until the
POWER
key is pressed.
15.2 Backlight
Although the backlight may be turned on and off at any time by pressing the
BACKLIGHT
key, there are
several available initial settings (applicable upon powering on the tester). To access these settings, select
Backlight
from the menu. The display appears as follows:
BACKLIGHT
Off
On
*
Auto
Set Minutes
1
INCORRECT
PASSWORD
Reset password
Request code:
XXXX
AUTOMATIC SHUTOFF
*
Disabled
Enabled
Set Minutes
5
www.
.com
information@itm.com