4 TEST FUNCTION SETUP
This section provides configuration instructions for each function.
After pressing
menu
, the initial Test Function Setup screen appears as follows:
Label
Description
ESC
Exits Test Function Setup, reverts to Operating Mode
< –
Scrolls to the previous function
– >
Scrolls to the next function
ENTR
Selects the function, allowing it to be modified
When the parameters have been configured as desired and are ready to be saved, press
ESC
to exit
Test Function Setup. The screen will show
SAVE CHANGES?
Pressing
YES
will save the changes and
return to the Operating Mode. Pressing
NO
will not save the changes and return to Operating Mode.
Note: Changes can be made to an unlimited number of settings before saving.
4.1 Speed
Independent speeds can be set for both directions of movement.
Default setting: 10 RPM / 60°/s
Available settings:
TSTM-DC / TSTMH-DC: 0.01 - 15 RPM [0.05 - 90°/s]
TSTM-DC-2 / TSTMH-DC-2: 0.04 - 60 RPM [0.2 - 360°/s]
4.2 Auto return
With this function, the torque plate moves to a limit (torque set point, angle/degree limit, or various event
completions in the M7I), whichever occurs first, and stops. Then, the torque plate returns to the other limit
and stops. The test speed is dictated by the setting described in the previous section. The return speed is
locked to maximum speed.
S P E E D : 2 0 . 0 0
E S C < − − > E N T R
C W S P E E D : 1 0 . 7 3
E S C − + E N T R
Label
Description
+
Increments the speed setting. Holding down
+
will
increment at an increasingly faster rate.
–
Decrements the speed setting. Holding down
–
will
decrement at an increasingly faster rate.
ENTR
Returns to the Test Function Setup menu
ESC
Exits the function without saving changes