Model WT3-200 Wire Terminal Pull Tester
User’s Guide
10
When set points are enabled, the following indicators are shown to the left of the primary reading:
– the displayed value is greater than the upper
force limit (NO GO HIGH)
– the displayed value is between the limits (GO)
– the displayed value is less than the lower force
limit (NO GO LOW)
Set point indicators and outputs reference the displayed
reading, not necessarily the current live load.
8.2 Set Point Outputs Schematic
9 DATA MEMORY AND STATISTICS
The WT3-200 has storage capacity of 1,000 data points. Readings may be stored, viewed, and output to
an external device. Individual, or all, data points may be deleted. Statistics are calculated for the data
presently in memory.
To enable memory storage, select
DATA Key
from the menu, then scroll to
Memory Storage
and press
ENTER
. Then exit the menu. In the home screen, the data record number
0000
will appear below the
primary reading. Press
DATA
at any time to save the displayed reading. The record number will
increment each time
DATA
is pressed. If
DATA
is pressed when memory is full the message “MEMORY
FULL” will be flashed at the bottom of the display and a double audio tone will be sounded.
To view, edit, and output stored readings and statistics, select
Memory
from the menu. The screen
appears as follows:
9.1 View Data
All the saved data points may be viewed. The record number is displayed, along with the corresponding
value and presently set unit of measurement. Any readings may be deleted individually. To do so, scroll
MEMORY
View Data
View Statistics
Output Data
Output Statistics
Output Data & Stats
Clear All Data