Model WT3-200 Wire Terminal Pull Tester
User’s Guide
20
Attach weight fixtures (brackets, hooks, etc), as required. Do not yet attach any weights or apply
any calibration loads. Then press
ENTER
.
7. The display will appear as follows:
Optionally exercise the load cell several times (at full scale, if possible), then press
ENTER
.
8. After displaying “PLEASE WAIT…” the display will appear as follows:
9. The display will appear as follows:
Apply a weight equal to the full scale of the instrument, then press
ENTER
.
Remove the load, leave the fixtures in place, then press
ZERO
.
10. The display will appear as follows:
CALIBRATION
ENSURE NO LOAD
THEN PRESS ZERO
CALIBRATION
GAIN ADJUST
APPLY FULL-SCALE LOAD
200.0 lbF +/-20%
THEN PRESS ENTER
CALIBRATION
Attach necessary
weight fixtures.
THEN PRESS ENTER
CALIBRATION
Optionally exercise
load cell a few times.
THEN PRESS ENTER
CALIBRATION
APPLY LOAD
1 OF 5
ENTER LOAD:
40.0 LBF
THEN PRESS ENTER