Model WT3-200 Wire Terminal Pull Tester
User’s Guide
24
13.2 Backlight
Several initial settings are available upon powering on the tester. To access these settings, select
Backlight
from the menu. The display will appear as follows:
Select
Off
for the backlight to be off upon powering on the tester.
Select
On
for the backlight to be on upon powering on the tester.
Select
Auto
for the backlight to be on upon powering on the tester, but will shut off after a period of
inactivity (as defined in the
Automatic Shutoff
sub-section). The backlight will turn on again when activity
resumes. The length of time of inactivity is programmed in minutes via the
Set Minutes
parameter.
Available settings:
1-10
, in 1 minute increments.
Note:
If the AC adapter is plugged in, the tester will ignore these settings and keep the backlight on.
Selecting the
On
or
Off
setting in the
Backlight
menu will manually turn the backlight on or off.
13.3 LCD Contrast
The contrast of the display may be adjusted. Select
LCD Contrast
from the menu. The screen will appear
as follows:
Press
ENTER
to modify the contrast. Select a value from
0
to
25
,
25
producing the most contrast.
13.4 Beeps
Audible tones can be enabled for all key presses and alerts, such as overload, set point value reached,
etc. The Set Point alert can be configured to be either a momentary tone or a continuous tone (until the
load is restored to a value between the set points). To configure the functions for which audible tones will
apply, select
Beeps
from the menu. The screen will appear as follows:
BACKLIGHT
Off
On
*
Auto
Set Minutes
1
BEEPS
Keys
* Alerts
Set Points
* Momentary
Continuous
LCD CONTRAST
Set Contrast
10