Model WT3-200 Wire Terminal Pull Tester
User’s Guide
26
14 SPECIFICATIONS
14.1 General
Force Capacity:
200 x 0.1 lbF | 3200 x 2 ozF | 100 x 0.05 kgF | 1000 x 0.5 N | 1 x 0.0005 kN
Accuracy:
±0.2% of full scale
Wire diameter range:
AWG30 - AWG 3 [0.03 - 0.25 in (0.8 - 6.3 mm)]
Min. sample length:
7.50 in [191.0 mm], excluding termination
Max. elongation:
1.15 in [29.2 mm]
Sampling rate:
7,000 Hz
Power:
AC or rechargeable battery. Low battery indicator appears when battery level is low, and
tester powers off automatically when power reaches critical stage.
Battery life:
Backlight on:
up to 7 hours of continuous use
Backlight off:
up to 24 hours of continuous use
Outputs:
USB / RS-232:
Fully configurable up to 115,200 baud. Includes Tester Control Language
2 for full computer control.
Mitutoyo (Digimatic):
Serial BCD suitable for all Mitutoyo SPC-compatible devices.
Analog:
±1 VCD, ±0.25% of full scale at capacity,
General purpose:
Three open drain outputs, one input.
Set points:
Three open drain lines.
Safe overload:
150% of full scale (display shows “OVER” at 110% and above)
Weight:
16.0 lb [7.3 kg]
Included accessories:
Universal voltage AC adapter, battery, quick-start guide, USB cable, resource CD (USB
driver, MESUR
TM
Lite software, MESUR
TM
gauge DEMO software, and user’s guide), and
NIST-traceable certificate of calibration with data.
Warranty:
3 years (see individual statement for further details)
14.2 Factory Settings
Parameter Setting
Set points
Upper
Disabled (defaults to 80% of full scale when enabled)
Lower
Disabled (defaults to 40% of full scale when enabled)
Filters
Current
8
Displayed
1024
DATA Key Functions
RS-232/USB Output
Enabled
Mitutoyo Output
Disabled
Memory Storage
Enabled
Backlight Auto
Minutes
1
Serial/USB
RS-232 Output Selected
Enabled
USB Output Selected
Disabled
Baud Rate
9,600
Data Format
N units
Mitutoyo BCD Output
Disabled
Break Detection
Automatic Output