16
17
B. To Recall Setting from Memory
1. Press
[ ]
key, all digits (Fig.1) and software version (Fig.2) will be shown on
LCD successively, then “bMI” will be displayed on LCD (Fig.13). Press
[ ]
key twice and “mem” will be displayed on LCD (Fig.30).
1. Press [ ] key, all digits (Fig.1) and software version (Fig.2) will be shown on LCD
successively, then “M” with last memorised location will be blinking (Fig.26).
2. Press [▲] or [▼] or numeric key 0-24 to select memory location. Press [ ] to confirm.
3. If the memory location is empty, “----” will display (Fig.27). Press [▲] / [▼] or input
numeric key 0-24 to select another memory location.
4. If the memory location is occupied, the ID No. and memorised personal data of BMI
parameters (height) or BFA parameters (age, gender, normal/athlete, height) and the
last BMI or BFA results will be shown on LCD, sequentially three times. Then it will
turn off automatically.
REPLAY MEASURED DATA FROM MEMORY
C
IMPORTANT NOTICE TO USERS
● The screen will show "---.-" if no results or unsuccessful estimation is recorded in the
memory.
2. Press [ ] to confirm and it will enter into memory recall mode. “M” with last memorised
location will blink (Fig.31). Press [▲] or [▼] or numeric keys 0-24 to select memory
location. Press [ ] to confirm the memory location.
3. If the memory is empty, “----” will be displayed (Fig.32). Press [▲] or [▼] or numeric
keys 0-24 to select another memory location.
4. If the memory is occupied, memorised ID No. and BFA parameters (height, age, gender,
normal or athlete selection) will be toggled on screen. Press [ ] to confirm and “o” will
cycle on screen.
5. Follow steps 6-8 in the section of
BODY FAT AND TOTAL BODY WATER ESTIMATING
OPERATION.
PROBLEM-SOLVING
1. No weight display?
Check if the scale is powered up and started up. If not, please refer to the Section
“
PREPARATION BEFORE USE
”.
2. "bAtt" is displayed on screen.
Battery is running low. Replace the battery.
3. The message displays “Err 0”….
The zero range is too great. Check if the platform is empty. If not, please remove
the weight off the platform.
4. The message displays “Err 2” when measuring?
Overload warning. Remove the weight immediately; otherwise, permanent damage
to the scale may occur.
5. Unable to hold weight.
Check whether your weight is stable. If not, please wait until “STA” icon is shown
on screen, and then press [
] again.
6. The message displays “Err 3” when estimating body fat and body water
values?
Contact error. Impedance cannot be measured. Please make sure that you are
standing still on the scale and maintain maximum contact between your feet and
the metal contacts. Please refer to Section “
Operation
”. You may need to moisten
your feet to improve the electrical contact.
7. The message displays “Err 4” when estimating body fat and body water
values?
Estimated body fat percentage is beyond the technical limit. Moistening of the feet
may help to improve the electrical contact.
8. The message displays “Err 5” when estimating body fat and body water
values?
Percentage total body water is out of range. Moistening of the feet may help to
improve the electrical contact.
9. The message displays “Err6”…
Machine error. Please contact your supplier.
10.The message displays “---.-” when replay...
Unsuccessful BMI or BFA operation. Follow related BMI measuring or BFA
estimating operation to complete your measurement.
11.The message displays “uStb” when measuring?
Unstable zero. Step off the scale, then step on the scale again.
12. After following corrective actions the problem still occurs.
If problem persists after following the corrective actions, disconnect the power
supply and reconnect after 1 minute.
13.The message displays “Adc”?
Machine error. Please contact your supplier.
14.The message displays “EEP”?
Machine error. Please contact your supplier.
15.The message displays “N.A”?
Without height rod. Check whether you have installed the height rod or the rod
signal cable has connected with RJ11 port. If not, Install height rod according to
related session in
Electronic Height Measuring Rod Instruction Manual
.
16.I have tried all corrective actions, but still can’t solve the problem …
Please contact your supplier.
C
IMPORTANT NOTICE TO USERS
● Do not step on the scale until "0.0kg" with "STA" indication shows on screen.