Current Type
HF Power
Crest Factor
HF Voltage
Pulsed Argon: slow repetition rate*
1–100 W at 500 ohms (CF 18–5.4) max. 4,000 V
p
max. 5,900 V
pp
Pulsed Argon: fast repetition rate *
1–100 W at 500 ohms (CF 18–5.4) max. 4,000 V
p
max.
5,900
V
pp
Bipo Pure Cut
1–120 W at 300 ohms
CF 1.7
max. 600 V
p
max. 1,200 V
pp
Bipo Blend Cut
1–120 W at 300 ohms
CF 2.1
max. 850 V
p
max.
1,700
V
pp
Forfex
1–120 W at 50 ohms
CF 1.9
max. 180 V
p
max.
360
V
pp
Micro Coag
0.1–50 W at 200 ohms
CF 1.8
max. 200 V
p
max.
400
V
pp
Micro Stop
0.1–50 W at 200 ohms
CF 1.8
max. 200 V
p
max. 400 V
pp
Auto Micro
0.1–50 W at 200 ohms
CF 1.8
max. 200 V
p
max.
400
V
pp
Macro Coag
1–120 W at 50 ohms
CF 1.9
max. 180 V
p
max. 360 V
pp
Macro Stop
1–120 W at 50 ohms
CF 1.9
max. 180 V
p
max.
360
V
pp
Auto Macro
1–120 W at 50 ohms
CF 1.9
max. 180 V
p
max. 360 V
pp
Seal Safe
5–320 W at 25 ohms
CF 1.9
max. 300 V
p
max.
600
V
pp
Endo Seal
1–100 W at 50 ohms
CF 2.1
max. 230 V
p
max. 450 V
pp
Note:
V
p
= peak voltage;
V
pp
= peak to peak
* Data applies to coagulation phase only; cycle-based average value deviates.
** Data averaged across cycle; pulse-based average value deviates.
n.a. = not applicable
Only the following currents are permitted for use with the baby NE (maximum power indicated in each case):
Pure Cut 100 W
Contact Coag 90 W
Argon Beam 100 W
Clamp Coag 90 W
Blend Cut 100 W
Contact Coag Auto Stop
90 W
Argon Endo 100 W
Forced Clamp 90 W
Super Blend 100 W
Spray 80 W
Care Coag 50 W
Pulsed Argon: slow
repetition rate 100 W
Argon Cut 100 W
Endo Spray 100 W
Care Coag Auto Stop 50 W
Pulsed Argon: fast
repetition rate 100 W
Care Cut 70 W