4. MAINTENANCE
4.1 Battery Replacement
To avoid shock or injury, disconnect the multimeter from any
external circuits or components and remove the test leads
before proceeding.
Replace the battery by removing the two screws from the battery
cover and lifting the battery cover.
Fit a new 9V, NEDA 1604, JIS006P, IEC 6F22 battery and replace
the battery cover and screws.
4.2 Fuse Replacement
To avoid shock, injury or damage to the multimeter,
disconnect the multimeter from any external circuits or
components and remove the test leads before proceeding.
Replace only with the fuse specified.
Replace the fuses by removing the three screws securing the rear
casing, and lift off the rear casing.
Replace F1 or F2 only with the original type 0.25A/500V, fast
acting ceramic fuse.
Replace the rear casing and screws.
4.3 Test Lead Replacement
If the test leads become damaged they should be replaced.
The replacement test leads must have the same (or better)
overvoltage category rating as the TL16 leads supplied
9