ELECTRICAL
All specified accuracies are at 23°C ± 5°C,
<75% RH for 1 year.
Temperature coefficient: 0.1 x (specified
accuracy) per °C (0°C to 18°C, 28°C to 50°C)
DC VOLTAGE
Ranges: 200mV, 2000mV, 20V, 200V, 600V
Resolution: 200mV, 2000mV, 20V, 200V, 600V
Accuracy: ± (0.5% of rdg + 1 dgt)
Input impedance: 10M
Ω
Overload protection: 600V DC or AC rms
600V DC or AC rms for 15 seconds on 200 mV
range
AC VOLTAGE (50Hz to 500Hz)
Ranges: 200mV, 2000mV, 20V, 200V, 600V
Resolution: 0.1mV, 0.001V, 0.01V, 0.1V, 1V
Accuracy: 200mV to 20V ranges ± (1.2% of rdg + 4 dgts)
200V to 600V range ± (2.0% of rdg + 4 dgts)
Input impedance: 10M
Ω
Overload protection: 600V DC or AC rms. 600V DC or AC rms for 15
seconds on 200mV range
DC CURRENT
Ranges: 200μA, 20mA, 200mA
Resolution: 0.1μA, 0.01mA, 0.1mA
Accuracy: ± (1.0% of rdg + 1 dgt)
Input protection: 0.25A/500V fast blow ceramic fuse
Specification
MM64
Digital Multimeter