1.1.2 INTERNAL ELECTRONIC FEATURES
The electronics include a voltage supervisor in charge of verifying the power
supply limits of the system and includes also a real-time clock. Samples are
taken at programmed time intervals. Every sample’s record is always
accompanied by a time-stamp. The logger incorporates a soft-start for external
power supplies. It also includes a pulse consumption circuitry for internal and
external battery discharge measurement
1
.
There is a
minimum gap of time between samples of 2 seconds
that should
be respected
.
2 seconds is also the resolution when defining a time interval, i.e.,
LOGS50 can be configured to make data acquisitions for instance, on
“15/03/2009, 17:15:(00 / 15 / 30 / 45)”
.
The logger has awakening capability by digital input transition.
1.1.3 INTERNAL POWER SUPPLY
The device has internal lithium battery cells supplying the necessary power for
its autonomous operation.
1
It will be the final user who decides, by means of the Dataloggers’ Configurator software CONF50 in the administrative
data file, over which part of the electronic circuitry the consumption pulse has to be done. This consumption pulse can
be done or over the internal battery or over the external battery included in the battery kit. The user will be able to
choose from these two options.