LAB-4222H-eIFU
Attaching the LNOP MP Patient Cable to the Masimo Sensor
1. Refer to
Fig. 2
. Orient the connecting tab of the sensor so that the «shiny» contacts are pointed up. Mate the logo on
the sensor to the logo on the patient cable.
2. Refer to
Fig. 3
. Insert the tab of the sensor into the patient cable connector until there is a tactile or audible click of
connection. Gently tug on the patient cable connector to ensure a positive contact. Tape may be used to secure the
cable to the patient for ease of movement.
Disconnecting the LNOP MP Patient Cable from the Masimo Sensor
1. Refer to
Fig. 4
. Place thumb and index finger on the buttons on either side of the patient cable connector.
2. Press firmly on the buttons and pull to remove the sensor.
SPECIFICATIONS
LNOP MP Patient Cables are for use with the following devices and sensors:
When used with Masimo SET® Technology:
Sensors
Body Weight
Saturation Accuracy
(70-100% SpO
2
)
Pulse Rate Accuracy
(25-240 bpm)
No Motion
Motion
Low
Perfusion
No Motion
Motion
Low
Perfusion
LNOP Adt/Adtx
> 30 kg
2%
3%
2%
3 bpm
5 bpm
3 bpm
LNOP Pdt/Pdtx
10-50 kg
2%
3%
2%
3 bpm
5 bpm
3 bpm
LNOP Inf-L
3-20 kg
2%
3%
2%
3 bpm
5 bpm
3 bpm
LNOP Neo-L
< 3 kg
3%
3%
3%
3 bpm
5 bpm
3 bpm
> 40 kg
2%
3%
2%
3 bpm
5 bpm
3 bpm
LNOP NeoPt-L
< 1 kg
3%
3%
3%
3 bpm
5 bpm
3 bpm
LNOP Trauma
> 30 kg
2%
3%
N/A
3 bpm
5 bpm
N/A
LNOP Newborn
Neonatal
< 3 kg
3%
3%
3%
3 bpm
5 bpm
3 bpm
LNOP Newborn Infant/
Pediatric
3-10 kg
2%
3%
2%
3 bpm
5 bpm
3 bpm
10-30 kg
LNOP Blue
2.5–30 kg
3%
(for 80–100)
N/A
3%
3 bpm
N/A
3 bpm
4%
(for 60–80)
3.3%
(for 70–100)
LNOP DC-195
> 30 kg
2%
3%
2%
3 bpm
5 bpm
3 bpm
LNOP YI
1-3 kg
3%
3%
3%
3 bpm
5 bpm
3 bpm
> 3 kg
2%
3%
2%
3 bpm
5 bpm
3 bpm
LNOP TC-I
> 30 kg
3.5%
N/A
3.5%
3 bpm
N/A
3 bpm
LNOP TF-I®
> 30 kg
2%
N/A
N/A
3 bpm
N/A
N/A