4523G-eIFU-0516

en 

rainbow® Series Patient Cables

  

For Pulse Oximeters with rainbow Technology

DIRECTIONS FOR USE

Reusable

LATEX

PCX-2108A

02/13

  Not made with natural rubber latex

Non-Sterile

INDICATIONS

The rainbow® Series Patient Cable Series have the same indications for use as the applicable sensors. Please see the 

compatible sensor’s directions for use for indications for use and prescribed information.

DESCRIPTION

The rainbow Series Patient Cables are used with Masimo® rainbow and M-LNCS™ Sensors.rainbow Series Patient Cables 

are for use only with devices containing Masimo rainbow SET™ technology Version 7.1 or higher or licensed to use 

rainbow compatible sensors. Consult individual oximetry system manufacturers for compatibility of particular sensor 

and device models. Each device manufacturer is responsible for determining whether their devices are compatible with 

each sensor model.
rainbow Series Patient Cables were verified using Masimo rainbow SET technology.

WARNING:

 Masimo sensors and cables are designed for use with devices containing Masimo SET® or Masimo rainbow 

SET oximetry or licensed to use Masimo sensors.

WARNINGS, CAUTIONS, AND NOTES

•  Always refer to the oximeter module operator’s manual for complete instructions or additional instructions. 

•  Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the cable 

and discard if cracks or discoloration are found.

•  All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor, cable 

and sensor before use, otherwise degraded performance and/or patient injury can result.

•  Carefully route patient cable to reduce the possibility of patient entanglement or strangulation.

•  Failure to properly connect the sensor or the oximeter module to the cable will result in intermittent readings, 

inaccurate results, or no reading.

•  To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or 

disconnecting either end.

•  To prevent damage, do not soak or immerse the sensor or cable in any liquid solution. Do not attempt to sterilize 

the cable or sensor.

•  Do not attempt to reprocess, recondition or recycle Masimo sensors or patient cables as these processes may 

damage the electrical components, potentially leading to patient harm.

•  Caution: 

Replace the patient cable when a Replace cable or a persistent low SIQ message is displayed  when the 

patient monitoring time is exhausted.  

•  Note: 

The patient cable is provided with X-Cal™ technology to minimize the risk of inaccurate readings and 

unanticipated loss of patient monitoring. The cable will provide 17,520 hours of patient monitoring time. Replace 

the cable when the patient monitoring time is exhausted.

CABLE AND SENSOR SELECTION

NOTE: The maximum length for the sensor, cable and extension cable combination is 15 feet. 
Refer to the table below for correct examples:

Cable 

Sensor 

Length

18 inches

3 feet

6 feet

1 ft

Yes

Yes

Yes

4 ft

Yes

Yes

No

12 ft

No

No

No