75
76
Table 2:
Terminals
1) Insulation resistance / voltage test: L(+), E(-), GUARD (GUARD terminals
can only be used for insulation resistance / leak current tests).
2) Temperature sensor, USB, and charger.
Max. power consumption
1) 10 VA (using battery);
2) 6 VA (using rechargeable battery pack).
Max. input voltage
750 VAC, ± 1000 VDC
Max. input frequency
70Hz
Max. rated voltage to earth
1000 Vrms (CAT III), 600 Vrms (CAT IV)
Insulation strength
6880 VAC: 15 seconds
Applicable standards
The design is complied with DL/T 845.1-2004
EN61010-1:2001
Overload protection
1000 VAC, 1200 VDC between L(+) and E(-) terminals: 1 minute
Dimensions
About 284 mm (L)×230 mm (w)×125 mm (h)
Net Weight
About 2.5 kg (without batteries)
Power supply
1) LR14 alkaline battery × 6; rated voltage: 1.5 V × 6
2) Battery pack, rechargeable nickel - metal hydride batteries; rated voltage:
7.2 V (for 500 times of charging)
Max. powering time
Alkaline battery: about 9 hours; Battery pack: about 5 hours (with
5 kV-voltage generated, with terminals in an open circuit, and with backlight
turned off).
Table 3:
Standard accessories
1. Test cable, 3 m × 3
2. Crocodile clips, ×3
3. Instruction manual, ×1
4. LR14 alkaline battery, ×6
5. USB cable, ×1
6. Temperature sensor, ×1
7. CD, ×1
Optional accessories
1. Rechargeable battery pack
2. AC charger
Temperature compensation function
Interface
USB Ver 2.0, for communications with the PC software.
PC application software
1: Data transfer
2: Tester setup
3: Data output
Table 4:
PI / DAR display function
Step-voltage measurement function
Data saving function: manual recording (100 records), log recording (10), deleting a single record, deleting all records,
uploading data to PC
Temperature/humidity input function: (input range for temperature, -10.0 to 70 °C; for humidity, 0.0 to 99.9% RH).
Timer function: (time can be chosen from 30 seconds to 30 minutes)
Display of the time spent for test
Time display: (including year, month, day, hour, minute, and second)