background image

4-2. Electrical Specification 
 

Function 

Range 

Best 

Resolution 

Accuracy 

Resistance 

300

Ω

/3K

Ω

30K

Ω

/300K/3M

Ω

 

 

30M

Ω

 

0.1

Ω

 

±

(1%rdg+ 2dgt) 

 

 

±

(1.2%rdg+ 3dgt) 

 

Capacitance 

3nF/30nF/300nF/ 

3uF/30uF/ 

300uF/3mF/30mF 

1pF 

±

(2%rdg+ 3dgt) 

 

±

(3%rdg+ 3dgt) 

Diode Check 

Open Voltage    :    2.8V 

Testing Current :    2mA 

 

Continuity 

Check 

When the resistance is less than 30

Ω

, the Buzzer is sounded. 

 

Accessory: 

        1. Spare Test Pin:        2 pcs 
 
 
               

 

       
 

5. Testing Operation 

  5-1, Scanning Mode 

When  install  the 3V  Lithium  Battery,  the  Tester  immediately  power  ON. 

In this case, the 

‘SCAN

’ and ‘

 - - - -

 ’ sign is displayed on the LCD. This 

is indicate the Tester is inter to the auto scanning mode and automatically 

identified the resistor, capacitor, diode and continuity 

and then enter to   

the corresponding measurement mode. 

 

In the Scanning Mode, when two pins are in touch with the object being 

measured, the measured value will be displayed on the LCD.   

 

Caution 

When measuring SMD device on the PCB, You must be 

Disconnect power and discharge all high- voltage   

Capacitors. 

5-2. Resistance Measurement 

      [1] 

  Auto scanning Mode & Auto range: 

                        300.0

Ω

 

  3.000M

Ω

 

      [2] Selected Resistance measurement Mode by the

FUNC

  key: 

            The range is extended to 30 M

Ω

, i.e.300.0

Ω

 

  30.00M

Ω

 

          When over load, the ‘OL’ symbol will be displayed on the LCD. 

5-3. Capacitance Measurement 

      [1] Auto scanning & Auto range: 3.000nF 

  300.0uF

 

      [2] Selected Capacitance Mode by the ‘FUNC’ key:   

            The range is extended to 30.00mF, i.e. 3.000nF

30.00mF. 

 

Summary of Contents for MS8910

Page 1: ...nd Indicators 3 1 Product Illustration 3 2 Push Buttons 3 3 LCD display 4 Specification 4 1 General Feature 4 2 Electrical Specification 5 Testing Operation 5 1 Scanning Mode 5 2 Resistance Measurement 5 3 Capacitance Measurement 5 4 Diode Check 5 5 Continuity Check 6 Maintenance 6 1 Replacing The Battery 6 2 Cleaning ...

Page 2: ... thermo plastic elastomer and the two testing pins has been gold plated so that it is reduce pin contact resistance and prevent to get rusty This operating instruction covers information on safety and caution so please read relevant information carefully and observe all the warming and note strictly 2 Safety Note The following safety notes should be observed prior to using This Tester and any asso...

Page 3: ...rm the Scanning Mode and then the SCAN and sign is displayed on the LCD This is indicate the Tester is inter to the auto scanning mode and will be automatically identify the resistor capacitor diode and continuity measurement mode The FUNC Button is used to selecting some one measurement mode manually 2 HOLD Press HOLD Button to enter or exit the hold mode In any measurement mode This key is act w...

Page 4: ...ion by one FUNC Push Button Data Hold function Continuity Checking function Over Load Indication OL Low Battery Indication Power Supply 3V Lithium Battery CR2032 1pc Test Pins is Gold Plated Auto Power OFF If the Tester is idle for more than 10 minutes The Tester automatically turns the power off Operating temperature Humidity 0 40 32 104 80 RH Storage temperature Humidity 10 50 14 122 70 RH Safet...

Page 5: ...canning mode and automatically identified the resistor capacitor diode and continuity and then enter to the corresponding measurement mode In the Scanning Mode when two pins are in touch with the object being measured the measured value will be displayed on the LCD Caution When measuring SMD device on the PCB You must be Disconnect power and discharge all high voltage Capacitors 5 2 Resistance Mea...

Page 6: ...n touch with on the component cathode The measured value show on the LCD Reverse the test pin and measure the voltage across the diode again If diode is good the display shows OL If diode is shorted the display shows 0 zero in both direction If display shows OL in both direction the diode is open 5 5 Continuity Check May selecting Scanning Mode or Continuity Mode by FUNC push button When the resis...

Reviews: