IEC 60601-1-2 Table 6
–
Input DC power port
Phenomenon
Basic EMC standard
IMMUNITY TEST LEVELS
Professional healthcare facility environment
HOME HEALTHCARE ENVIRONMENT
Electrical fast
transients
/bursts
IEC 61000-4-4
+/-2 kV
100 kHz repetition frequency
Surges
Line-to-line
IEC 61000-4-5
+/-0,5 kV, +/-1 kV
Surges
Line-to-ground
IEC 61000-4-5
+/- 0,5 kV, +/- 1 kV, +/- 2 kV
Conducted
disturbances
induced by RF
fields
IEC 61000-4-6
3 V
0,15 MHz
–
80 MHz
6 V in ISM bands
between 0,15 MHz
and 80 MHz
80 % AM at 1 kHz
3 V
0,15 MHz
–
80 MHz
6 V in ISM and amateur radio
bands between 0,15 MHz and 80
MHz
80 % AM at 1 kHz
Electrical transient
conduction along
supply lines
ISO 7637-2
Not applicable
As specified in ISO 7637-2