MATOS ARIA Eco 150L User Manual v.02 24/04/2018
Electrical Connection
The appliance is intended for connection to alternating current. The connection values for voltage (V)
and frequency (Hz) are declared on the name plate in the cabinet. Power must be connected via a wall
socket with a switch. It is strongly recommended that the wall socket is easily accessible.
All earthing requirements stipulated by the local electricity authority. The cabinet plug and wall socket
should give correct earthing. If in doubt, contact your local supplier or an authorised electrician.
If the supply cord is damaged, it must be replaced by the manufacturer, its service agent or similarly
qualified persons in order to avoid a hazard.
WARNING
: THIS APPLIANCE MUST BE EARTHED
The flexible cord (mains lead) fitted to this appliance has three cores for use with a 3-pin 13 amp plug.
If a BS 1363 (13-amp) fused plug is used it should be fitted with a 13-amp fuse.
The cores in the mains lead are coloured in accordance with the following code:
GREEN AND YELLOW: EARTH
BLUE: NEUTRAL
BROWN: LIVE
These colours might not correspond with the colour markings identifying the terminals in your plug.
Note:
•
First Connection to the mains
•
The appliance provides information about the higher temperature (Hi alarm) till the moment
of cooling.
•
The alarm is signalised by a sound signal and “Hi” caption on the display. To cancel the
alarm, press any button.
•
Until the moment of cooling, the alarm is repeated every 15 minutes. Such an operating mode
is normal until the required temperature settings are reached.
Green & Yellow
13 Amp Fuse
Brown
Blue
Cord clamp