INTENDED USE

The intended use of the Matrix MD-S70 Medical Leg Press is to create a 

predetermined, defined physical activity level for the patient. The physical activity 

level is generated by the user pushing against a stationary foot plate to move the 

seat against a chosen resistance. Alternatively, both the foot plate and the seat 

may be locked into stationary positions, allowing the user to perform isometric 

exercises. This physical activity level can be done in the controlled environment 

of a doctor’s office or a clinician’s lab. The stress intensity of the physical activity 

level can be adjusted by adjusting the amount of resistance applied to the chair, 

or by the perceived effort as described by the user to the doctor or clinician.

Metabolic System: The medical leg press can be used to influence the metabolic 

system to increase fat-burning, thereby reducing the size of adipose tissue deposits.

Physical Performance Tests: The defined stress intensity of the 

physical activity level can be used to perform physical performance 

tests on the strength of the user’s leg muscles. 

Recovery: Under a doctor’s care, the medical leg press can be used to improve 

muscle strength after surgery or injury. The handrails and seat of the medical 

leg press can be used for additional safety and stabilization of the patient.

Orthopedic Rehabilitation: Under a doctor’s care, the medical leg 

press can be used for orthopedic rehabilitation (e.g. supporting 

defined/reduced weight to strengthen the legs of a user).

ONE LEG

1

2

DECLINE POSITION

1

2

BOTH LEGS

1

2

PROPER USAGE

1.  Do not exceed weight limits of the exercise device.
2.  If applicable, set safety stops to appropriate height.
3.  If applicable, adjust seat pads, leg pads, foot pads, range of motion 

adjustment, or any other type of adjustment mechanisms to a comfortable 

start position. Make certain that the adjusting mechanism is fully 

engaged to prevent unintentional movement and to avoid injury.

4.  Sit on bench (if applicable) and get into appropriate position for exercise.
5.  Exercise using no more weight than you can safely lift and control.
6.  In a controlled manner, perform exercise.
7.  Return weight to its fully-supported start position.

Summary of Contents for MD-S70

Page 1: ...MD S70 MEDICAL LEG PRESS ...

Page 2: ...2 ...

Page 3: ... avoid injury INSTALLATION 1 STABLE AND LEVEL SURFACE MATRIX exercise equipment must be installed on a stable base and properly leveled 2 SECURING EQUIPMENT Manufacturer recommends that all stationary MATRIX strength equipment be secured to the floor to stabilize equipment and eliminate rocking or tipping over This must be performed by a licensed contractor 3 Under no circumstances should you slid...

Page 4: ...ements 4 PLYOMETRIC Explosive movement training or jump training How to choose your elastic band strength and wrap the band or bands through the blue band pegs on the outside of the frame Move the weight stack disengagement lever to the unlocked position Perform your plyometric exercise as desired CAUTION Take note of the range limiting device location Manufacturer suggests positioning the limitin...

Page 5: ...ty level can be used to perform physical performance tests on the strength of the user s leg muscles Recovery Under a doctor s care the medical leg press can be used to improve muscle strength after surgery or injury The handrails and seat of the medical leg press can be used for additional safety and stabilization of the patient Orthopedic Rehabilitation Under a doctor s care the medical leg pres...

Page 6: ...sition platform and then release Allows the Therapist to adjust the starting position of the footplate The therapist steps down on the adjustment pad slides the platform closer or further away from the user to achieve the desired knee flexion at the start position and then releases the adjustment pad It is easier to do when standing at the rear of the unit versus the side 3 Foot Rest Grasp center ...

Page 7: ...tion release pin Allows the user to adjust the starting position of the back pad The user pulls the pin and slides the back pad closer or further away to achieve a comfortable start position 8 Range Limiter Pull pin and slide carriage bumper forward or backward to limit the carriage range of motion ADJUSTMENT POINTS 7 Carriage Lock To lock the carriage into a fixed position grasp handle and rotate...

Page 8: ...AINTENANCE CHECKLIST ACTION FREQUENCY Clean Upholstery 1 Daily Inspect Cables 2 Daily Clean Guide Rods Monthly Inspect Hardware Monthly Inspect Frame Bi Annually Clean Machine As Needed Clean Grips 1 As Needed Lubricate Guide Rods 3 As Needed 1 Upholstery Grips should be cleaned with a mild soap and water or a non ammonia based cleaner 2 Cables should be inspected for cracks or frays and immediate...

Page 9: ...or to assembly confirm all the components by matching them with the exploded diagrams Carefully unpack the unit from this box and dispose of the packing materials in accordance with your local laws Exercise device must be assembled by a trained assembly team or be pre assembled prior to delivery CAUTION To avoid injury to yourself and prevent damage to the frame components be sure to have proper a...

Page 10: ...her Φ10 2 Nut M10 Hex Head Bolt M10x20L Spring Washer Φ10 2 Washer Φ10 2 4 8 4 2 2 2 Do not fully tighten frame connectors until assembly is complete Vibra Tite 135 Red Gel or equivalent must be used on all fasteners that are not assembled with Nylock Nuts Keep pull pin in the lock position ENGLISH ...

Page 11: ...ty D E F Hex Head Bolt M10x20L Spring Washer Φ10 2 Washer Φ10 2 2 2 2 Step 4 Hardware Description Qty B D E F G H Arc Washer Φ10 2 Hex Head Bolt M10x20L Spring Washer Φ10 2 Washer Φ10 2 Socket Head Bolt M10x20L Pre installed Nut M12 2 2 4 2 2 1 Cable ENGLISH ...

Page 12: ...12 5 6 K F J I F F C Step 6 Hardware Description Qty C F I J K Nut M10 Washer Φ10 2 Bolt M10x75L Bolt M10x30L Pre installed Screw M5x20L 6 14 6 2 4 ENGLISH ...

Page 13: ...13 2 1 3 F G 7 8 Step 7 Hardware Description Qty F G Washer Φ10 2 Socket Head Bolt M10x20L 3 3 Route cable through hole and attach using pre installed hardware ENGLISH ...

Page 14: ... 9 Hardware Description Qty K L M Pre installed Screw M5x20L Screw M8x20L Washer Φ8 2 4 4 4 10 F G N Step 10 Hardware Description Qty F G N Washer Φ10 2 Socket Head Bolt M10x20L Socket Head Bolt M10x25L 4 4 2 ENGLISH ...

Page 15: ...15 O N P Q N O 3 1 2 11 12 Step 12 Hardware Description Qty N O P Q Socket Head Bolt M10x25L Plastic Cap Bolt M8x15L Rubber Stopper 2 2 2 2 ENGLISH ...

Page 16: ...cription Qty D F G Hex Head Bolt M10x20L Washer Φ10 2 Socket Head Bolt M10x20L 2 4 2 14 S T R U Step 14 Hardware Description Qty R S T U Special Screw Nylon Washer Φ13 2 Plastic Spacer Ring 4 4 4 4 Torque Value 2 5 Nm 22 in lbs ENGLISH ...

Page 17: ...17 V M F 15 16 Step 15 Hardware Description Qty F Washer Φ10 2 2 Step 16 Hardware Description Qty M V Washer Φ8 2 Bolt M8x25L 4 4 ASSEMBLY COMPLETE ...

Page 18: ...2 2 2 INCREMENTAL WEIGHTS Set of 4 with holder Each weight adds 0 9 kg 2 lbs of effective resistance to the exercise Used to attach a Tyrmotion TYMO force plate sold separately to the front of the Leg Press foot plate Small platform used to help users enter the machine Contains the following items attached magnetically to the front of the Leg Press foot plate Heel Cup Wedge Block Wire Mount holds ...

Page 19: ...19 ...

Page 20: ...MD S70 2019 Johnson Health Tech Part XXXXXXXXXX Rev 1 2 A ...

Reviews: