7
11. Setting Waveform
Three types of waveform can be selected: sine wave, triangular wave and square wave.
The options of the advance option mode vary with the waveform selected.
12. Protection Functions
Type
Details
Setting screen
Remarks
Voltage limit
protection
(V limit)
The output is
limited to the
setting.
The setting value is
expressed as only a
positive value, but
protection also
takes place in the
negative side.
Current limit
protection
(I limit)
Overvoltage
protection
(OVP)
The output is cut off
at the DC voltage
setting.
After cutting off, the product is
put in FAULT state.
To exit FAULT state, press the
OUTPUT switch.
Overcurrent
protection
(OCP)
The output is cut off
at the DC current
setting.
Frequency setting range: 10 mH - 30 kHz
When the FREQ switch is pressed and held down,
the frequency setting value changes in the following
manner:
1Hz
→
5Hz
→
10Hz
→
50Hz
→
100Hz
→
500Hz
→
1kHz
→
5kHz
→
10kHz
→
20kHz
→
10mHz
→
50mHz
→
100mHz
→
500mHz
→
1Hz
Sine wave
Select Sine.
Advanced
options of the
sine waveform
Triangular wave
Select Triangle.
Advanced options
of the triangular
waveform
Square wave
Select SQUARE.
Advanced options
of the square
waveform
Ex.) Voltage limit operation
- voltage
(voltage limit
setting)
+ voltage
OFF: Normal state (Output OFF)
ON: Normal state (Output ON)
ON: Fault state (Output OFF)