7 Synchronized operation option
102
DOEF series
Matsusada Precision
7-2-4 Input and Output Signals of Synchronous Clock
[Input signal]
Voltage: 0.5 to 5Vp-p
Input impedance: 1kΩ (AC-coupling)
Input frequency: 10MHz (+/- 0.5%)
[Output signal]
Voltage: 1Vp-p (AC-
coupling, 50Ω (typical)
Output impedance: 50Ω (AC -coupling)
Output frequency: 10MHz
✽
The common for both input and output signals is isolated from the chassis. The voltage to ground
is up to 42 Vpk (peak).
-
Unit power up on the synchronous trigger operation:
For synchronous trigger operation, turn on the power of both the synchronous master and the
synchronous slave units at the same time, or turn on the synchronous master unit first, and
then turn on the synchronous slave units.
-
Unit power up on the synchronous clock operation:
For synchronous clock operation, turn on the reference unit and the units receiving the clock
at the same time, or turn on from the reference unit to units receiving the clock in order.
If units receiving the clock are turned on before the reference unit, the clock may not be
locked properly.