5 Sequence Function
73
DOEF series
Matsusada Precision
5-4-9 Program Parameters
Program parameters are a setting that determines the behavior of a program. Such program
parameters and other settings are shown in the following table.
Table: Program parameters and other settings
Parameter
Description
Setting value
End step
The step finally executed in a program.
1 to Maximum step
No. of loops
The repeat count. It is counted up when the last
step is executed.
0 to 9999 and infinity
End setting
Sets the behavior of the output after the last step
in the final loop was executed.
OUTPUT OFF
Hold (OUTPUT ON)
Program linking operation
No. of programs
The program count.
When it is modified, all sequence data will be
initialized.
1 to 64
Initialization of all
program data
All sequence data in CV or CC mode will be
initialized to factory default.
The value of the maximum step varies depending on the number of programs.
Fig. Each settings in program edit
1. End step
Select
“End Step” on the program edit screen using the downward
“▼”
or upward
“▲” movement
switch. Set your desired value by turning the DIAL.
2. No. of loops
Select
“Loop Number” on the program edit screen using the downward
“▼
” or upward “▲”
movement switch. Set your desired value by turning the DIAL.
When
“0” is set, the steps are not repeated. When “---“ is set, the number of loops is set to
infinite.
Prog No.
:
3
End Step
:
32
Loop Number
:
100
End Setting
:
OFF
Change Prog Num
64
Program Data Clear
End step
Loop count
End setting
Change of program count
Initialization
Summary of Contents for DOEF Series
Page 1: ...Instruction Manual F RA 001 3R2 MODEL DOEF series B N 243 9 024 243 9 024 Rev 0 4 ...
Page 15: ...2 External Appearance 7 DOEF series Matsusada Precision 2 3 Dimensions 2 3 1 400W Model ...
Page 16: ...2 External Appearance 8 DOEF series Matsusada Precision 2 3 2 600W Model ...
Page 17: ...2 External Appearance 9 DOEF series Matsusada Precision 2 3 3 1200W Model ...
Page 121: ......
Page 123: ......
Page 124: ......