1-4 Caution for handling
1.
Make the setting of an output voltage to zero (0). Turn off the HV ON/OFF switch.
2.
Make sure that the output voltage is zero with the output voltmeter of this unit.
3.
Turn off the POWER ON/OFF switch.
4.
Ground the output for more than 10 seconds, and make sure that the output voltage is zero
at another HV voltmeter. It is especially dangerous when a capacitive load is connected or a
long cable is attached there to.
5.
Make it a rule to touch loads with right hand.
・For safer operation, be sure to ground the ground terminal of power supply at
One point on the ground.
・
Make sure to connect the GROUND terminal correctly as shown above. Improper grounding
is dangerous and may cause an electric shock or damage the power supply.
・
In case output short circuit or discharging is expected, use shorter and thicker grounding
wire.
WHEN TOUCHING LOAD AFTER TURNING OFF A HIGH VOLTAGE
Connection for bi-polar type
1 Introduction
Wires should be :
・more than 5SQ
・less than about 2m
・not bared
!
MUST
Ground at one point
!
MUST
Output cable
CH2
CH1
+ + + + +
- - - - -
chassis
Ground terminal
Electrode
In case not connected to chassis,
make sure to ground at one point
!
MUST