5-5 Command Syntax
For example, the following programming sets the output voltage of the power supply connected UNIT1 to
1000V:
①
Enter a delimiter.
・
RS-232C/RS-485 connection
CR(0D
H
)
・
LAN connection
CR(0D
H
)
・
USB connection
CR(0D
H
)
・
GPIB connection
GPIB delimiter
(Refer to instruction manual of CO series)
②
Some commands require a parameter.
③
Wrong commands are ignored.
④
Specify the UNIT number.
Use #0 to #31 or #AL.
NOTE
The number of characters for a command shall be less than
20
, including a delimiter.
5-6 Multi-command
Specifying "AL" for the UNIT number causes the command to be sent to all units connected.
(Format examples)
①
#AL VSET 1000
Set the output voltage of all devices to 1000V.
②
#AL SW1
Enable the output of all units.
(Remarks) Refer to the command description for commands that support multi-command execution.
5. Option
#1
VSET
1000
○○
②
parameter
③
command
④
UNIT No.
①
delimiter