11
REN Command
Enable remote control.
Format
examples
REN
Initial
value
GTL
Remarks
・
When this command is received, all commands can be used.
・
When this command is received, “RE” is displayed on the screen.
This command supports multi-command execution.
GTL Command
Disable remote control.
Format
examples
GTL
Initial
value
GTL
Remarks
・
When this command is received, all commands other than REN, STS, and
output measurement are ignored.
・
When this command is received, the “RE” display on the screen turns off.
This command supports multi-command execution.
Switching between local control and remote control is also possible with the “LCL / RMT switch” on
the front panel.
VGET Command
Output voltage monitor.
Operating
range
0 to Rated voltage
resolution
1
unit
V
Format
examples
VGET
Return
examples
VGET=1234
⇒
output 1234V
Remarks
IGET Command
Output current monitor.
Operating
range
0 to rated current
resolution
0.001
unit
mA
Format
examples
IGET
Return
examples
IGET=1.234
⇒
output 1.234mA
Remarks
5. Option
Summary of Contents for EP Series
Page 1: ...Rev 0 2 B N 035 9 114 035 9 114 MODEL EP series Instruction Manual ...
Page 19: ......