4 Basic Operation and Screens
22
HARS series
4-5-3 Overvoltage protection
When the Output voltage becomes higher than the set OVERVOLTAGE value, OVP is activated
and the output will be cut off approx. 0.1 seconds later after its detection.
When OVP is activated, "OVER VOLTAGE" appears on the touch panel display and the output will
be cut off.
In order to clear the state, touch the [Enter] button on the touch panel display or set the external
switch to OFF. (The "OVER VOLTAGE" display will disappear and return to the normal screen.)
To restart the output, adjust the output voltage or OVER VOLTAGE setting value, and then set the
ON/OFF switch to "ON" or the External Switch to "ON".
4-5-4 Overcurrent protection
When the Output current becomes higher than the set OVER CURRENT value, OCP is
activated and the output will be cut off approx. 0.1 seconds later after its detection.
When OCP is activated, "OVER CURRENT" appears on the touch panel display and the output
will be cut off.
In order to clear the state, touch the [Enter] button on the touch panel display or set the external
switch to OFF. (The "OVER CURRENT" display will disappear and return to the normal screen.)
To restart the output, adjust the output current or OVER CURRENT setting value, and then set the
ON/OFF switch to "ON" or the External Switch to "ON".
4-5-5 Interlock function
When the shorted the interlock remains open,
“INTERLOCK” appears on the touch panel
display, and the output wil be cut off.
When the interlock function is activated, "INTERLOCK" appears on the touch panel display and
the output will be cut off.
To clear the state, the interlock is shortened, and then, touch the [Enter] button on the touch panel
display. (The "INTERLOCK" display will disappear and return to the normal screen.)
To restart the output, set the ON/OFF switch to "ON" or the External Switch to "ON".
Summary of Contents for HARS Series
Page 1: ...Instruction Manual F RA 001 3R3 MODEL HARS series B N 277 9 002 277 9 002 Rev 0 1 ...
Page 10: ......
Page 19: ...2 Before Using This Product 9 HARS series 2 6 3 Dimensions a Models 500W 1kW 60kV or lower ...
Page 20: ...2 Before Using This Product 10 HARS series b Models 500W 1kW 100kV 120kV ...
Page 21: ...2 Before Using This Product 11 HARS series c Models 2kW 3kW 60kV or lower ...
Page 22: ...2 Before Using This Product 12 HARS series d Models 2kW 3kW 100kV 120kV ...
Page 92: ......