5 Operation on the Main Unit
29
HARS series
Arc count (2)
5-1-9-2
The output is turned off when a set number of arcs have occurred within a set time after the arc has
occurred.
The arc count shows the number of times the arc has occurred within the set time. By touching the
Arc count reset button, the arc count can be reset to "0".
To enable the function of turning off the output according to the arc count, touch the Arc count button
and set it to "Enable".
When the arc protect count setting value is touched, the protect count value for the arc for which the
output is to be turned off can be set with the setting change dial.
Touch the "Arc protection enable time" to set the enable time of the arc count to turn off the output
with the setting change dial.
When the arc count exceeds the arc protect count setting value, the output is turned off.
Setting screen
Home screen
Arc count
Arc count button
Arc total count
Arc count reset button
Setting value of Arc protect
count
Arc protection enable time
Arc count (2) setting screen
Summary of Contents for HARS Series
Page 1: ...Instruction Manual F RA 001 3R3 MODEL HARS series B N 277 9 002 277 9 002 Rev 0 1 ...
Page 10: ......
Page 19: ...2 Before Using This Product 9 HARS series 2 6 3 Dimensions a Models 500W 1kW 60kV or lower ...
Page 20: ...2 Before Using This Product 10 HARS series b Models 500W 1kW 100kV 120kV ...
Page 21: ...2 Before Using This Product 11 HARS series c Models 2kW 3kW 60kV or lower ...
Page 22: ...2 Before Using This Product 12 HARS series d Models 2kW 3kW 100kV 120kV ...
Page 92: ......