5-1-8 Configuring Multiset Memory
To change each set value, touch the respective item on the touch panel display.
To enable the multiset, touch the Multiset ON/OFF button
to set “Enable”.
By touching the Memory number button, the output setting value will change to the multiset number
you selected.
* As of the limit (voltage limit/current limit), when each setting (voltage/current) value is input, the
input value is limited by the value set in H-Limit.
* If each setting (voltage/current) exceeds the configuration set in H-Lmit while the multiset function is
enabled, the setting value is limited to the H-Limit value.
* Note that the output remains on even if the multiset memory number is changed while the output is
ON.
* When each setting (voltage/current) value is input while the output is OFF, the status LED lights up
in blue
Setting screen
Home screen
Multiset memory configuring screen
Multiset ON/OFF button
Memory number
button
Voltage
Current
Memory-A
Output setting value
Memory-B
Output setting value
Memory-C
Output setting value