6 Operating by Analog Remote Control
49
HARS series
6-3 Using Interlock
You can turn the output Force ON/OFF using Output Switch. This function can be implemented according
to the following method.
Set the interlock switch to ON/OFF (short = output enabled, open = output OFF*).
If the switch is open, the touch panel display will show an "INTERLOCK" error message.
When shorted, output is enabled by the OUTPUT ON/OFF switch or External Switch.
*When the interlock function is disabled, the interlock protection function is not available.
TTL LOW
=
OUTPUT:ON enable
TTL HIGH
=
OUTPUT:OFF
or
External relay SHORT
=
OUTPUT:ON enable
External relay OPEN
=
OUTPUT:OFF
13
25
14
1
TB1
TTL signal
External relay
External connection
+
-
8
17
13
25
14
1
Internal circuit
of unit
470Ω
Summary of Contents for HARS Series
Page 1: ...Instruction Manual F RA 001 3R3 MODEL HARS series B N 277 9 002 277 9 002 Rev 0 1 ...
Page 10: ......
Page 19: ...2 Before Using This Product 9 HARS series 2 6 3 Dimensions a Models 500W 1kW 60kV or lower ...
Page 20: ...2 Before Using This Product 10 HARS series b Models 500W 1kW 100kV 120kV ...
Page 21: ...2 Before Using This Product 11 HARS series c Models 2kW 3kW 60kV or lower ...
Page 22: ...2 Before Using This Product 12 HARS series d Models 2kW 3kW 100kV 120kV ...
Page 92: ......