7 Operating with Digital Control
57
HARS series
7-4 Description of SCPI Commands
*IDN?
Gets company name, model name, serial No., version info.
Format
*IDN?
Response ex.
Matsusada Precision,HARS-60P16.7, 123456S, Ver.01.00.00
([company name],[model name],[serial No.],[version info])
*RST
Resets parameters.
Format
*RST
Reset details
Turn off the output.
Resets the settings of output voltage, output current, and each protection to the
factory default.
SYSTem:LOCal
Sets local mode.
Format
SYST: LOC
Initial value
SYST:LOC
Remarks
When this command is received, commands except for "*IDN?", "*RST",
"SYST", "MEASure", "SYSTem:MODe?" are all ignored.
SYST:REMote
Sets remote mode.
Format
SYST: REM
Initial value
SYST:LOC
Remarks
When this command is received, all commands will become available.
SYSTem:MODe?
Gets the unit status.
Format
SYST: MODe?
Query
CF CC OVP LD
[Output state]
␣
[Operational mode]
␣
[Error status]
Remarks
If there are multiple errors, all of them are returned.
INSTruments [:NSELect] <n | ALL>
Sets valid unit number for command.
Range
0 to 31 | ALL
Unit
INST 31
Format
INST ALL
Query
Command returns the valid UNIT number for INST?.
Summary of Contents for HARS Series
Page 1: ...Instruction Manual F RA 001 3R3 MODEL HARS series B N 277 9 002 277 9 002 Rev 0 1 ...
Page 10: ......
Page 19: ...2 Before Using This Product 9 HARS series 2 6 3 Dimensions a Models 500W 1kW 60kV or lower ...
Page 20: ...2 Before Using This Product 10 HARS series b Models 500W 1kW 100kV 120kV ...
Page 21: ...2 Before Using This Product 11 HARS series c Models 2kW 3kW 60kV or lower ...
Page 22: ...2 Before Using This Product 12 HARS series d Models 2kW 3kW 100kV 120kV ...
Page 92: ......