32
MBW973-SF6_MANUAL_E_V3.1
Once the screw cover has been removed, the black optical
assembly (optical head) is now removed by pulling it straight
toward you. The mirror is now visible and you are ready to
perform the Ice Test.
Confirm that you opened the measuring head and are ready
for the Ice Test by pressing the
Ok
button.
Note:
The test starts immediately after pressing the
Ok
button.
During the Ice Test, the mirror rapidly cools to approximately
-30 °C. Because the measuring head is open, humidity from
the ambient air starts to condense on the mirror. This forms
a frost layer on the mirror which can be strengthened if nec-
essary by blowing on it. After reaching the low temperature
and forming ice on its surface, the mirror begins heating. As
the temperature approaches 0 °C, the instrument begins to
beep, with those beeps becoming faster as the mirror gets
closer to ice-melting temperature.
Visually observe the mirror. As soon as the mirror tempera-
ture crosses 0 °C, the ice will melt into liquid water drops
(phase transition).
When you observe the phase transition on the mirror, press
the
Ok
button. The mirror temperature is measured at that
moment and a dialog box appears with the test results.
Summary of Contents for 973-SF6