Aqueous solutions of glycol are corrossive to metals and therefore it must be used with an
inhibitor. This corrossion inhibitor forms a surface barrier that protects the metal from attack.
The inhibited glycol solution are generally stable and relatively non-corrossive to most standard
materials. The exception is galvinised steel because the zinc coating will react with the inhibitor
part found in most formulated inhibited glycol sold commercially.
Performance effects of qlycol solutions:
a. Higher concentrations of glycol will retard heat transfer coefficient in the BPHE. For
example, a 30% glycol concentration may cause a drop of capacity by about 20%. Refer to
the attached graph depicting this drop of performance (Appendix 16).
b. As the concentration increases, the viscosity of the solution also increases. Therefore, the
friction losses of the piping system will also increase. A correction factor must be used to
evaluate this increase. See Appendix 16.
c. Similarly, the viscosity increase will also affect the water pump performance. The flow rate and
efficiency will reduce with higher concentrations. Thus, the head pressure developed will also
drop. See also Appendix 16.
Due consideration of these effects must be given when designing and selecting the chiller unit
for such a system. Failure to do so may result in insufficient capacity and water flow rate.
Guidelines:
1. Before applying the glycol solutions, thoroughly clean and flush the system. Reaction with
sludge, rust, deposits and oil may retard the inhibitor function. This is especially so after using
cleaning agents in the BPHE. Complete removal of the cleaning agent is necessary before
charging in the glycol.
2. Calculate the total water volume in the system and determine the amount of glycol needed.
The solution can be mixed outside the system in drums or barrels and then pumped into the
system.
Use clean and soft water (low in chloride and sulfate ions) for the mixing. Distilled water,
deionized water or condensate water may be used for this purpose.
3. The system must not be connected to an automatic make-up water pipeline. This may
cause dilution of the concentration and increases the water freezing point.
The inhibitor portion in the solution may also deplete over time due to reaction with
contaminants in the system.
Therefore, it is vital that a maintenance programme is in place to determine the glycol
concentration in the system. This can be measured by using refractometer (which measures
the refractive index of the solution), gas chromatography or density bottle (though not so
accurate).
4. Do not mix different inhibited glycol formulations together.
Section 9
Page 114
Summary of Contents for AC 40A
Page 1: ...AM MAC ...
Page 153: ...Appendix 1 ...
Page 154: ...Appendix 2 Appendix 2 ...
Page 157: ...GRAPH OF PRESSURE DROP VS FLOW FOR VARIOUS C V RATINGS Appendix 5 1 ...
Page 158: ......
Page 159: ...Water Capacity Graph Appendix 5 2 ...
Page 160: ...CHART 3 FRICTION LOSS FOR CLOSED PIPING SYSTEMS Appendix 6 1 ...
Page 161: ...CHART 4 FRICTION LOSS FOR OPEN PIPING SYSTEMS Appendix 6 2 ...
Page 162: ...CHART 5 FRICTION LOSS FOR CLOSED AND OPEN PIPING SYSTEMS Appendix 6 3 ...
Page 168: ...Appendix 8 Appendix 8 ...
Page 169: ...Appendix 8 1 ...
Page 170: ...Appendix 8 2 ...
Page 171: ...Appendix 8 3 ...
Page 172: ...Appendix 8 4 ...
Page 173: ...Appendix 8 5 Appendix 8 5 ...
Page 174: ...Appendix 8 6 Appendix 8 6 ...
Page 175: ...Appendix 8 7 Appendix 8 7 ...
Page 176: ...Appendix 9 ...
Page 178: ...Appendix 11 1 Appendix 11 1 ...
Page 179: ...Appendix 11 2 Appendix 11 2 ...
Page 180: ...Appendix 13 ...
Page 181: ...Appendix 13 1 ...
Page 182: ...Appendix 14 Appendix 14 ...
Page 183: ...Appendix 15 1 Appendix 15 1 ...
Page 186: ...Appendix 17 Appendix 17 ...
Page 187: ...Appendix 18 ...
Page 188: ...CE20DW CE25DW Note PRESSURE DROP CORRECTION FACTOR 1 2947 0 0021 EWTºF Appendix 19 ...
Page 189: ... 2003 McQuay International 603 55194922 www mcquayup com ISO 9002 REGISTERED ...