background image

Top hat rail housing GSV-1H

The measuring amplifier GSV-1H is available with voltage outputs ±5V, ±10V and current 
output 4...20mA.

The force sensor is connected to terminals 1 to 4.
The voltage supply (11-28V) is connected to terminal 5 and 8 (ground).

For zero setting, terminal 6 is connected to the mains supply or a control signal between 5 
and 24 volts is applied.

The amplification can be switched 1-2-4-10-times with internal jumpers.

If the input signal is negative, output follows up to 0mA the input signal, beyond it remains 
at 0mA.

Pin Configuration

Pins
1 / 2

+U

S

U

S

Bridge supply (for sensor input, "excitation“, "input“)

3 / 4

+U

D

U

D

Differential input (for sensor output “output“, sensor signal)

5 / 6

+U

B

Tara

DC voltage supply

+12V or +24V

Steuereingang Nullabgleich

7 / 8

U

A

GND

Analog output

Masse

Note: The ground of the voltage supply is connected to terminal 8.

Dimensions

Instruction Manual

7

GSV-1H:
Depth: 114.5mm
for installation on mounting rails according to 
EN50022

Summary of Contents for GSV-1

Page 1: ...Straingage Bridge Amplifier GSV 1 Version July 09 2007 ME Me systeme GmbH Neuendorftstr 18a Tel 49 3302 559 282 DE 16761Hennigsdorf Fax 49 3302 559 141...

Page 2: ...iagram 10 Pin Configuration 11 Adjustment of Input Sensitivity 11 Layout of Jumpers 12 Dimensions 12 Option EP cConnection of an External Potentiometer 12 Aluminum housing GSV 1A 14 Pin Configuration...

Page 3: ...Strain Gage Measuring Amplifier GSV 1 Instruction Manual 3...

Page 4: ...ply up to 4 parallel connected weighing cells with 350 ohm bridge resistance each and is therefore eminently suitable for applications in weighing technology The option GSV H can supply for 8 weighing...

Page 5: ...ification stages 1x 2x 4x 10x and trimmer 1x 10x 5V 10V 4 20mA 0 20mA GSV 1H amplification switching Option TR stepless amplification setting via trimmer potentiometer HSW Version Top hat rail housing...

Page 6: ...ce optional for GSV 1H 5 6 7 5 47 10V 4 20mA 0 20mA V V Ohm Operating voltage Nominal range Operating range Nominal range GSV 1A GSV 1A 12 0 24 10 5 28 GSV 1L 10 5 16 or 18 28 V V V Current consumptio...

Page 7: ...can be switched 1 2 4 10 times with internal jumpers If the input signal is negative output follows up to 0mA the input signal beyond it remains at 0mA Pin Configuration Pins 1 2 US US Bridge supply...

Page 8: ...t voltage at sensitivity option 2 mV V Output voltage at sensitivity option 3 5 mV V 1 1 5 volts at 2 mV V 5 volts at 3 5 mV V 2 2 5 volts at 1 mV V 10 volts at 2 mV V 5 volts at 1 75 mV V 10 volts at...

Page 9: ...rol signal between 5 volts and 24 volts is fed in In this way the zero point of the force sensor will be balanced and the peak value storage will be reset to zero Through internal jumpers the amplific...

Page 10: ...an be selected Mode A actual value mode Signal sequence Amplification peak value storage The threshold indicator reacts to the actual value The actual value is shown on terminal 6 The peak value is sh...

Page 11: ...d of the supply voltage and the ground of the output signal are connected to terminal 8 Adjustment of Input Sensitivity The input sensitivity can be adapted by displacement of the jumper The input sen...

Page 12: ...he ground 12 Instruction Manual 99 0 22 5 1 2 3 4 5 6 7 8 91 01 11 2 1 31 41 51 6 GSV 1HSW SW 1 SW2 SW 1 SW2 Depth 114 5mm for mounting on carrying rails according to EN50022 On exceeding the threshol...

Page 13: ...tput 0 10V is realised with the amplification level 2 b In the standard design the device is equipped with a Bessel filter 2nd order 250 Hz c The calibration by the manufacturer is carried out either...

Page 14: ...o setting terminal 8 is connected to the voltage supply or a control signal between 5 volts and 24 volts is applied Amplification can be switched 1 2 4 10 times by means of jumpers Pin Configuration T...

Page 15: ...factor Output voltage at sensitivity option 2 mV V Output voltage at sensitivity option 3 5 mV V 1 1 5 volts at 2 mV V 5 volts at 3 5 mV V 2 2 5 volts at 1 mV V 10 volts at 2 mV V 5 volts at 1 75 mV...

Page 16: ...Option 718 plug The measuring amplifier GSV 1A is also supplied with plugs of type type 718 The pin configuration is the same as that of module housing GSV 1M 16 Instruction Manual...

Page 17: ...put are connected to the round plug B pin contacts For zero setting Pin 4 is connected to the voltage supply Pin 1 or a control signal between 5 Volt and 24 Volt is applied Dimensions Pin Configuratio...

Page 18: ...oltage supply 2 UD positive differential input 2 GND ground 3 US positive bridge supply 3 5V stabilized 4 US negative bridge supply GND 4 internally reserved 5 GND ground 5 internally reserved 6 UA an...

Page 19: ......

Page 20: ...en Alle Angaben beschreiben unsere Produkte in allgemeiner Form Sie stellen keine Eigenschaftszusicherung im Sinne des 459 Abs 2 BGB dar und begr nden keine Haftung Made in Germany Copyright 1999 2007...

Reviews: