Page 31
Control Commands
optoNCDT ILR 1181 / 1182
7.2.3
DW......Distance Tracking with Cooperative Target (10 Hz)
The command is only relevant for the ILR 1181.
Essential input parameters:
SA, SD, SE, SF, OF
Effect on:
RS232/RS422, digital switching output, analog output
- Performs at a steady measuring rate of 10 Hz.
- Stable measuring values only with a white target board at the target.
- No sudden jumps in distance > 16 cm.
7.2.4
DX......Distance Tracking with Cooperative Target
The command is only relevant for the ILR 1182.
Essential input parameters:
SA, SD, SE, SF, OF
Effect on:
RS232/RS422, digital switching output, analog output
- Performs at a steady measuring rate of 50 Hz.
- Stable measuring values only with a white target board at the target.
- Homogeneous motions with maximum 4 m/s.
- High rate of measurement, included preceding measuring values in the process to calculate a currently
measuring value.
- No sudden jumps in distance >16 cm.
7.2.5
DF......Distance Measurement with External Trigger
Essential input parameters:
SD, SE, SF, ST, OF, TD
Effect on:
RS232/RS422, digital switching output, analog output
- Preparation for the single measurement, triggered by an external trigger pulse.
Initially, after selecting this mode, the operator does not receive any response. As soon as the trigger pulse
has been detected, the sensor will send data and switches the digital and/or the analog output.
The Settings for the trigger delay (delay) and the trigger flank can be defined via parameter TD, see Chap.
7.3.13
.