Control Commands
optoNCDT ILR 1181 / 1182
7.3.16 AS....Display/Set Autostart Command
Standard setting: ID
AS (autostart) defines which function will be carried out when power becomes available to the sensor. Pos-
sible entries are those delivering a measuring value on the output side, an ID command and the command
for turning the laser on (LO).
For example, if ASDT has been parameterized, the sensor will begin with distance tracking on turning on
power.
Possible versions: DT/DS/DW/DX/DF/DM/TP/LO/ID
7.3.17 OFx.x.....Display/Set Distance Offset
Standard setting: 0
With the help of OF (offset) define a zero-point for his/her application. For details on the position of the mod-
ule’s zero-point, see Chap.
5.
- OF must be selected so it is properly matched to the currently valid scale factor setting (SF).
- OF may also take on negative values.
7.3.18 SO......Set Current Distance to Offset
SO performs a distance measurement and saves the measured reading as an offset value with inverted math-
ematical sign (OF).
Result: (offset = - distance)