9.
Online Help Tool
Once communication has been established with a PC (as described above), an online help tool can be called
up by triggering an ID [Enter] or id [Enter] command at the keypad. Its purpose is to support work with
distance measurement and parameterization commands. [Enter] corresponds to hexadecimal 0D
hex
(carriage
return).
DT[Enter] distancetracking
DS[Enter]
distancetracking 7 m
DW[Enter]
distancetracking with cooperetive target (10 Hz)
DX[Enter]
distancetracking with cooperetive target (50 Hz)
DF[Enter]
distance measurement with external trigger
DM[Enter]
distance measurement
TP[Enter]
internal temperature [°C]
SA[Enter] / SAx[Enter]
display/set average value [1..20]
SD[Enter] / SDd[Enter]
display/set display format [d/h]
ST[Enter] / STx[Enter]
display/set measure time [0..25]
SF[Enter] / SFx.x[Enter]
display/set scale factor
SE[Enter] / SEx[Enter]
display/set error mode [0/1/2]
0..Iout=const., ALARM=const.
1..Iout: 3 mA @RE>RB, 21 mA @RE<RB,
ALARM: OFF@AH>0, ON@AH<0
2..Iout: 21 mA @RE>RB, 3mA @RE<RB,
ALARM: ON@AH>0, OFF@AH<0
AC[Enter] / ACx.x[Enter]
display/set ALARM center
AH[Enter] / AHx.x[Enter]
display/set ALARM hysterese
AW[Enter] / AWx.x[Enter]
display/set ALARM width
RB[Enter] / RBx.x[Enter]
display/set distance of Iout=4 mA
RE[Enter] / REx.x[Enter]
display/set distance of Iout=20 mA
RM[Enter] / RMx y.y z[Enter] remove measurement
TD[Enter] / TDx y[Enter]
display/set trigger delay [0..9999 ms] trigger level [0/1]
TM[Enter] / TMx y[Enter]
display/set trigger mode [0/1] trigger level [0/1]
BR[Enter] / BRx[Enter]
display/set baud rate [2400..38400]