background image

CARING FOR YOUR TELESCOPE

Your telescope is a precision optical instrument designed for a lifetime of 

rewarding viewing. It will rarely, if ever, require factory servicing or mainte-

nance. Follow these guidelines to keep your telescope in the best condition:

Avoid cleaning the telescope’s optics. A little dust on the mirror or lens sur-

faces of the telescope will not cause loss of image quality. When absolutely 

necessary, dust on the lens or mirrors should be removed with very gentle 

strokes of a camel hair brush or blown off with an ear syringe (available at 

most pharmacies).

Fingerprints and organic materials on the optics may be removed with a 

solution of 3 parts distilled water to 1 part isopropyl alcohol. You may also 

add 1 drop of biodegradable dish washing soap per pint of solution. Use 

soft, white facial tissues and make short, gentle strokes. Change tissues 

often. 

CAUTION:

 Do not use scented or lotioned tissues or damage could 

result to the optics. DO NOT use a commercial photographic lens cleaner.

COLLIMATION (ALIGNMENT) OF OPTICS

All Meade reflecting telescopes are optically aligned at the factory prior to 

shipment. It is unlikely that you will need to align, or collimate, the optics 

after receipt of the instrument. However, if the telescope received unusu-

ally rough handling in shipment, it is possible that the optics must be re 

aligned for best optical performance. In any case this alignment procedure 

is simple, and requires only a few minutes the very first time the telescope 

is used. Take the time to familiarize yourself with the following collimation 

procedure, so that you will recognize a properly collimated instrument and 

can adjust the collimation yourself, if necessary.

1. CORRECT COLLIMATION

The properly collimated (aligned) mirror system in the Meade telescope 

assures the sharpest images possible. This occurs when the primary mir-

ror and secondary mirror are tilted so that the focused image falls directly 

through the center of the focuser draw tube. These mirror tilt adjustments 

are made with the secondary mirror assembly and the primary mirror cell 

(Fig. E,F,G) and will be discussed later. 

To inspect the view of the mirror collimation, look down the focuser draw 

tube with the eyepiece removed. The edge of the focuser drawtube (1, Fig. 

H), will frame the reflections of the primary mirror with the 3 mirror clips (2, 

Fig. H), the secondary mirror (3, Fig. H), the spider vanes (4, Fig. H), and 

your eye (5, Fig. H). Properly aligned, all of these reflections will appear 

concentric (i.e., centered) as illustrated in Fig. H. Any deviation from the 

concentric reflections will require adjustments to the secondary mirror as

-

sembly and/or the primary mirror cell (Fig. F,G).

2. SECONDARY MIRROR HOLDER ADJUSTMENTS

If the secondary mirror (1, Fig. I) is centered in the draw tube (2, Fig. I), but 

the primary mirror is only partially visible in the reflection (3, Fig. I), one or 

more of the 3 secondary mirror collimation screws need adjusting. First, un-

thread each of the secondary mirror collimation screws slightly to the point 

of where you can tilt the secondary holder from side-to-side. By grasping 

the secondary holder with your hand, tilt the secondary mirror holder until 

you see the primary mirror become as centered in the reflection of the di

-

agonal mirror as possible. Once you are at the best position, thread in the 

3 secondary mirror collimation screws to lock the holder in place. Then, if 

necessary, make adjustments to these 3 screws to refine the tilt-angle of 

the secondary mirror until the entire primary mirror can be seen centered 

within the secondary mirror reflection. When the secondary mirror is cor

-

rectly aligned, it will look like Fig. J (Note: The primary mirror is shown out 

of alignment).

3. PRIMARY MIRROR ADJUSTMENTS

If the secondary mirror (1, Fig. J) and the reflection of the primary mirror (2, 

Fig. J) appear centered within the draw tube (3, Fig. J), but the reflection 

of your eye and the reflection of the secondary mirror (4, Fig. J) appear off 

center, you will need to adjust the primary mirror tilt screws of the prima-

ry mirror cell (2, Fig. G). These primary tilt screws are located behind the 

primary mirror, at the lower end of the main tube. The long screws lock the 

mirror in place where the shorter screws perform the adjustment.

To adjust the primary mirror tilt screws (2, Fig G), first turn by several turns, 

the primary mirror cell lock knobs (3, Fig. G) that are next to each primary 

mirror tilt screw. The three primary mirror cell locking screws are Phillips 

head screws and are longer than the adjustment screws.

Then by trial-and-error, turn the short primary mirror tilt screws (2, Fig. G) 

until you develop a feel for which way to turn each knob to center the reflec

-

tion of your eye. Once centered, as in Fig. H, turn the 3 long primary mirror 

cell locking screws (3, Fig. G) lightly to re-lock the tilt-angle adjustment.

4. STAR TESTING THE COLLIMATION

With the collimation performed, you will want to test the accuracy of the 

alignment on a star. Use the 20mm eyepiece and point the telescope at 

a moderately bright (second or third magnitude) star, then center the star 

image  in  the  telescope’s  field-of-view.  With  the  star  centered  follow  the 

method below:

• Bring the star image slowly out of focus until one or more rings are visi

-

ble around the central disc. If the collimation was performed correctly, the 

central star disk and rings will be concentric circles, with a dark spot dead 

center within the out-of-focus star disk (this is the shadow of the secondary 

mirror), as shown in Fig. K3. (An improperly aligned telescope will reveal 

elongated circles (Fig. K1), with an off-center dark shadow.)

•If the out-of-focus star disk appears elongated (Fig. K1), you will need to 

adjust the primary mirror adjusting tilt screws of the primary mirror cell (3, 

Fig. G)

•To adjust the primary mirror tilt screws (3, Fig. G), first unscrew several 

turns the 3 hex-head primary mirror cell locking screws (2, Fig. G), to allow 

free turning movement of the tilt knobs.

•Using the flexible cable controls move the telescope until the star image is 

at the edge of the field-of-view in the eyepiece, as in Fig. K2.

•As you make adjustments to the primary mirror tilt screws (3, Fig. G), you 

will notice that the out-of-focus star disk image will move across the eye-

piece field. Choose one of the 3 primary mirror tilt screws and slightly move 

the shadow to the center of the disk. Then slightly move the telescope using 

the flexible cable controls to center the star disk image in the center of the 

eyepiece.

•  If  any  further  adjustments  are  necessary,  repeat  this  process  as  many 

times as needed until the out-of-focus star disk appears as in Fig. K3, when 

the star disk image is in the center of the eyepiece field.

• With the star testing of the collimation complete, tighten the 3 hex-head 

primary mirror locking screws (2, Fig. G)

CHANGING THE VIEWFINDER BATTERY

If the viewfinder red dot does not illuminate, verify the viewfinder is on by ro

-

tating the knob below the viewfinder lens clockwise. If the red dot does not 

illuminate, the battery may need replacing. To replace the battery, press the 

left side of the viewfinder housing labeled “push”. The battery compartment 

will slide out on the right side of the viewfinder (see Fig. L). Replace the 

battery with a Lithium CR-2032 battery with the positive side up. Then push 

the battery compartment back into the viewfinder and power on.

MEADE LIMITED WARRANTY

Every Meade telescope and telescope accessory is warranted by Meade 

Instruments Corp (“Meade”) to be free of defects in materials and work-

manship for a period of ONE YEAR from the date of original purchase in 

the U.S.A.  Meade will repair or replace a product, or part thereof, found by 

Meade to be defective, provided the defective part is  returned to Meade, 

freight-prepaid, with proof of purchase. This warranty applies to the original 

purchaser only and is non-transferable. Meade products purchased out-

side North America are not included in this warranty, but are covered under 

separate warranties issued by Meade international distributors. RGA Num-

ber Required: Prior to the return of any product or part, a Return Goods 

Authorization (RGA) number must be obtained from Meade by writing,  or 

by calling (800) 626-3233. Each returned part or product must include a 

written statement detailing the nature of  the claimed defect, as well as the 

owner’s name, address, and phone number. This warranty is not valid in 

cases where the product has been abused or mishandled, where unautho-

rized repairs have been attempted or performed, or where depreciation of 

the product is due to normal wear-and-tear. Meade specifically disclaims 

special, indirect, or consequential damages or lost profit which may result 

from a breach of this warranty. Any implied warranties which cannot be 

disclaimed are hereby limited to a term of one year from the date of original 

retail purchase. This warranty gives you specific rights. You may have other 

rights which vary from state to state. Meade reserves the right to change 

product specifications or to discontinue products without notice. This war

-

ranty supersedes all previous Meade product warranties.

VISIT MEADE.COM FOR MORE INFORMATION!

CR-2032 

battery

Battery

compartment

Figure L

Push here to 

open battery 

compartment

3

2

4

5

3

3

1

2

2

2

2

1

3

4

1

4

5

3

3

1

2

2

2

2

1

3

4

1

4

5

3

3

1

2

2

2

2

1

3

4

1

Figure E

Figure F

Figure I

Figure H

Figure G

Figure K

Figure J

1

3

2

Meade Instruments Corp. 27 Hubble, Irvine, CA. 92618, USA

                            

REV0. 20160606

Reviews: