Measurement Specialties
KPSI
®
Digital Series User’s Manual
Page 15
www.Meas-Spec.com
other concurrent measurements
The SDI standard allows for other concurrent measurement commands, such as “C1,” “C2,” etc.
The sensor supports the following optional concurrent measurements commands:
aC1!
Measure psi using factory calibration. Does not apply user scaling, field calibration,
or offsets. Returns 1 value and the units are fixed to psi.
aC2!
Measure temperature (Celsius or Fahrenheit). This returns two values: the
temperature and the units. The units will be 0 for Celsius and 1 for Fahrenheit.
aC3!
Measure user scale factors: user
slope
, user
offset
, field calibration
offset
. Use this if
you want to view user-entered values that can affect the value returned.
aC4!
Measure calibration lab slope and offset. Use this if you want to view the calibration
lab values that can affect the value returned.
aC5!
This command requests a concurrent high accuracy printed circuit board (PCB)
temperature.
aC6!
This command requests concurrent battery voltage.
aC7!
This command combines C0 and C2 for both pressure and temperature.
Remember to issue the “D0” command after the measurement is complete (after estimated time
for completion expires), in order to retrieve the data.