General Information 

 

Read and understand all operating instructions before using this product.

 

Keep this manual for future reference. 

 

Don’t use sharp objects to operate the keyboard. The waterproof  membrane may become damaged.   

 

Record the weight shortly after loading the platform. After extended periods, the load cell’s output signature may 

result in a less accurate reading. 

 

Place the scale on a hard, flat surface. Locate the leveling bubble and adjust the base feet as needed to ensure 

the scale is level before using. 

 

Avoid extended exposure to extreme heat or cold. Optimum operation is at normal room temperature. Allow the 

scale to acclimate to room temperature before using. 

 

Allow sufficient warm up time. Turn the scale on and allow up to 2 minutes for internal components to stabilize 

before weighing. 

 

Electronic scales are precision instruments. Do not operate near cell phones, radios, computers or other 

electronic devices that emit radio frequencies that may cause unstable readings. If the scale performs poorly, try 

moving to a different room or location.   

 

When storing the scale for extended periods, the battery must be charged every 90 days to avoid premature 

performance degradation. Over time, the operating time per charge will degrade. If the operating time is no 

longer acceptable, the battery must be replaced. 

 

Avoid using in heavy vibration or heavy airflow conditions. 

 

 

 
 

Specifications 

Model 

EH-SS-3 

EH-SS-6 

EH-SS-15 

EH-SS-30 

Max Capacity 

3 kg (6.6 lb) 

6 kg (13.2 lb)

15 kg (33 lb) 

30 kg (66 lb)

Readability 

0.5 g (0.001 lb)

1 g (0.002 lb)

2 g (0.005 lb) 

5 g (0.01lb)

Divisions 

6000 6000

7500 6000

Construction 

Washdown stainless steel housing, stainless steel pan 

Waterproof  Rating 

IP68

Weighing Units 

kg, g, lb, lb:oz, pcs

Modes 

Weighing, counting, checkweighing 

Working Temp 

25°F to 95°F (-5 to 35°C)

Storage Temp 

-10°F to 120°F (-25 to 50°C) 

Power 

Lead-acid rechargeable battery or AC power cord 

Max Tare 

100% Max capacity

Platform Dimensions 

230 x 190mm; 9” x 7.5”

Base Dimensions   

(L x W x H) 

287 x 236 x 110mm; 

11.3” x 9.3” x 4.3”

 
Contents   

 

Scale 

 

 

Operation manual 

 

Summary of Contents for EH-SS

Page 1: ...EH SS Operation Manual Contents subject to change without notice Version 1 0 7 2017 ...

Page 2: ......

Page 3: ...ic devices that emit radio frequencies that may cause unstable readings If the scale performs poorly try moving to a different room or location When storing the scale for extended periods the battery must be charged every 90 days to avoid premature performance degradation Over time the operating time per charge will degrade If the operating time is no longer acceptable the battery must be replaced...

Page 4: ...ill automatically turn off after 10 minutes with no key activity or weight change Note If the battery voltage falls below 5 6V the scale will automatically power off Setting the Tare Weight 1 Place an empty container on the platform and press the TARE key The display will return to zero eliminating the weight of the container The ZERO light will go off and the TARE light will be lit 2 Put the mate...

Page 5: ...er than lower limit If it is not oFF will be displayed and the scale will return to the normal weighing mode and checkweighing will be deactivated 6 Place objects on the scale platform The weight will be shown on the display If the weight is within the specified range the Accept light will be lit and an audible tone will sound If the value is outside the specified range the Over or Under light wil...

Page 6: ...ower off the scale 2 Turn the scale upside down remove the seal and use a small screwdriver to remove the two screws holding the calibration switch cover Move the calibration switch to the ON position toward the front of the scale 3 Place the scale on stable level surface and press the ON ZERO key to turn the scale on 4 After the self check is complete press and hold the TARE key until CAL is brie...

Page 7: ... PcS Counting mode piece count will be displayed weight will not Note Calibration must be done in Kg or g A oFF y Auto off mode n No auto power off y Auto off after 10 minutes of no operation or weight change FILtX FILt3 Display response speed FILt1 one step to display FILt2 3 4 steps to display FILt3 6 8 steps to display FILt4 display is fast ZEro 1d Set auto zero range 0 5d 1d 1 5d 2d 2 5d 3d 3 ...

Page 8: ...e will vary with use Over time the operating time per each full charge will degrade If the operating time is no longer acceptable the battery must be replaced When storing the scale for extended periods the battery must be charged every 90 days to avoid premature performance degradation Character Prompt Definitions dc x xx Battery voltage is x xx volts Scale capacity has been exceeded Reduce the w...

Reviews: