EHC-PF
.
15
.
“
UNLOAD
”(this means that the scale is ready to calibrate the zero point position, please remove any weight on the
scale)
,
the total price window display the output inner code of A/D. When the scale is stable and the unit indicator stops
flashing, press ENTER key to confirm the zero point calibration. After the scale is stable and gets the zero point, the scale
will go to the next step automatically.
6.11
The display of the weight window remains the same, the unit price window displays “
Load
”, which means the scale is
ready to calibrate the standard weight. The display of the total price and weight window remain the same. Place a
standard weight between 25%-100% FS on the center of the scale, press ENTER to confirm the standard weight
calibration after the scale is stable and the unit indicator stops flashing. When the scale gets the stable data, it will go to
the next step automatically.
6.12
The display of the weight window remains the same, the unit price window displays “
INPU.LD
”(Input Load Weight), the
total price window displays 0, use 0 - 9 keys or CLEAR to input loaded standard weight, then press ENTER key for
confirmation, the input data will be shown on the total price window.
6.13
When the weight window displays “
UNLOAD
” again
,
the scale is ready to re-confirm the zero point, remove any weight
on the scale, after the scale is ready and unit indicator stops flashing, press ENTER to confirm.
6.14
After calibration completes, UNIT PRICE window displays “
CAL.END
” and TOTAL PRICE window displays current weight.
6.15
If there’s an error occurred in calibration, the scale will display
CAL.ERR
; It usually means incorrect data input or
loading weight, and please try to return back to the last step to disposal.
6.16
Once the calibration switch is changed to OFF; the scale will exit the calibration mode.
7.
The meaning of some displayed symbols
Err01
:
Weight signal is too large.
Err02
:
No proper data can be displayed.
Err03
:
Weight signal is too small.
Err04
:
Zero point is over the setting range.
Err05
:
Zero point is below the setting range.
Err06
:
Error in unit key operation.
Err10
:
the EEPROM can’t be accessed.
Err11
:
The parameters are not same with backup data.
Err12
:
The setting parameter(s) is not in normal range.
Err20
:
There is an error in calibration.
Err30
:
ADC is over max. range.
Err31
:
ADC is below min. range.
Err40
:
Recall error, no content stored.
CAP.
:
Capacity.
UOL.
:
Voltage.
St.PLU
:
To set and store the indirect unit price.
ADDR.
:
PLU address.
Unit
:
Weighing unit selecting.
Rc.PLU
:
Recall indirect unit price.
UNLOAD
:
To unload the weight.
LOAD
:
To load the weight.