User Parameter Setup Mode 

1.

 

From normal weighing mode, press UNIT and ZERO/ON/OFF until “USer” is displayed. 

2.

 

Within user parameter setup mode: 

 

Press PRINT/HOLD to shift the flashing data entry position from right to left   

 

Press UNIT to increase the digit in the flashing data entry position by 1 

 

Press TARE to confirm and save the set data and enter next setting   

 

Press ZERO/ON/OFF key to exit this mode 

 

Parameter 

Default 

Setting 

U1 05 

Auto-off time:   

 

0 = No auto-off function     

 

01-15 = Scale will auto power off after 1-15 minutes with no activity 

U2 2 

Backlight on-off mode option: 

 

0 = Backlight is always off     

 

1 = Backlight is always on     

 

2 = Backlight is auto-off after 10 seconds with no activity   

U3 2 

PRINT/HOLD key function set: 

 

0 = Hold only     

 

1 = Print only     

 

2 = Hold and Print 

U4 0 

Hold function mode:

 

0 = No hold function     

 

1 = Holds the largest weight reading 

 

2-50 = Displays the stable weight within 3 seconds. 
Note: When weight exceeds 10d and the variation is within ±2d

±50d, the stable weight 

is displayed until less than 10d is detected on the platform. 
1d = 1 division = 1 unit of readability as noted in Specifications 

U5 1 

Serial communication output format: 

 

0= Communication is disabled 

 

1 = Outputs stable weight, unit, and status when PRINT/HOLD is pressed, but does not 
receive data 

 

2 = Output gross, tare, net weight, unit, and status when PRINT/HOLD is pressed, but does 
not receive data 

 

3 = Continuously outputs displayed weight, unit, and status, but does not receive data 

 

4 = Continuously outputs gross, tare, net weight, unit, and status, but does not receive data

 

5 = Outputs weight, unit, and status one time after scale is becomes stable   

 

6 = Outputs gross, tare, net weight, unit, and status one time after scale becomes stable 

 

7 = Two way communication mode 

U6 3 

Baud rate for Serial communication: 

 

0 = 1200 

 

3 = 9600 

 

1 = 2400 

 

4 = 19200 

 

2 = 4800 

U7 0 

Serial communication byte format:   

 

0 = 8N1 – 8 data bits, no parity, 1 stop bit 

 

1 = 7O1 – 7 data bits, 1 bit odd parity, 1 stop bit 

 

2 = 7E1 – 7 data bits, 1 bit even parity, 1 stop bit 

Summary of Contents for PS-102-200

Page 1: ...W W W M E A S U R E T E K N E T PS 102 Operation Manual Contents subject to change without notice Version 1 0 11 2011 ...

Page 2: ......

Page 3: ...vices that emit radio frequencies that may cause unstable readings If the scale performs poorly try moving to a different room or location Avoid using in heavy vibration or heavy airflow conditions Specifications Model PS 102 75 PS 102 200 Max Capacity 165lb 75kg 440lb 200kg Readability 0 1lb 0 05kg 0 2lb 0 1kg Resolution 1 1500 1 2000 Construction Epoxy painted carbon steel Stainless steel platfo...

Page 4: ...ition from right to left UNIT Weighing mode Chooses weight unit Setup Calibration mode Increases the digit in the flashing data entry position by 1 TARE Weighing mode Tares the weight Setup or Calibration mode Confirms the input data and forwards to next step ON OFF ZERO Weighing mode 3 seconds Zeros the platform weight 3 seconds Powers off the scale Setup or Calibration mode Exits to normal weigh...

Page 5: ... The default calibration input weight will be 50 of the scale s rated capacity Use the PRINT HOLD and UNIT keys adjust to the standard calibration weight you will use Note do not use less than 10 of the scale s rated capacity Place the standard calibration weight on the platform and press TARE to confirm your input and move to the next step If CAL Er is displayed press ZERO ON OFF key to exit this...

Page 6: ...ight within 3 seconds Note When weight exceeds 10d and the variation is within 2d 50d the stable weight is displayed until less than 10d is detected on the platform 1d 1 division 1 unit of readability as noted in Specifications U5 1 Serial communication output format 0 Communication is disabled 1 Outputs stable weight unit and status when PRINT HOLD is pressed but does not receive data 2 Output gr...

Page 7: ...et reading minus decimal point unit CR LF H1H2H3 CR ETX Bytes used 8 Weight reading 1 Minus 1 Decimal point 2 or 5 Weight unit 3 Current status H1 H2 H3 U5 3 Continuously outputs the current displayed reading weight unit and current status data but will not receive data The output format is the same as U5 1 U5 4 Continuously outputs the current gross weight tare weight net weight data weight unit ...

Page 8: ...haracter indicates positive weight The minus sign follows after the first digit Leading zero digits are suppressed Serial Communication Symbols LF Line Feed character hex 0AH CR Carriage Return character hex 0DH ETX End of Text character hex 03 SP Space hex 20H H1H2H3 Three status bytes see below for byte definitions p Polarity minus sign for negative weight and space for positive weight W1 W8 Wei...

Page 9: ...F H1H2H3 CR ETX Zero function is activated and returns to current scale status like pressing ZERO ON OFF If zero function cannot be activated it will return to current scale status Command T CR 54h 0dh Response LF H1H2H3 CR ETX Tare function is activated and returns to current scale status like pressing TARE If tare function cannot be activated it will return to current scale status Command U CR 5...

Page 10: ... 3 6V Replacement Parts Part Number Description MH12R97601G AC120V DC6V 500mA UL adaptor One Year Limited Warranty MeasureTek products covered in this manual are guaranteed to be free from defects in material and workmanship for a period of one year after date of purchase Misuse accidental damage overload alteration and improper installation are expressly excluded Any product which is determined t...

Reviews: