Calibration Mode 

Note: 

 

A standard calibration weight of  at least 10% of  the scale’s rated capacity is required. 

 

The calibration operation must be in pounds. 

 

Press ZERO/ON/OFF at any time to exit calibration mode. 

 

Using 80% to 100% of  the scale’s rated capacity is recommended for optimum results. 

 

1.

 

From normal weighing mode, press TARE and ZERO/ON/OFF until “CAL-?” is displayed. 

2.

 

Within calibration mode: 

 

Press PRINT/HOLD to shift the flashing data entry position from right to left   

 

Press UNIT to increase the digit in the flashing data entry position by 1 

 

Press TARE to confirm and save the set data and enter next setting   

 

Press ZERO/ON/OFF key to exit this mode 

3.

 

Press   TARE   to confirm and enter calibration mode.   

4.

 

“CAL.P0“will be displayed to calibrate scale’s zero-point. Remove all weight from the scale platform and press   TARE  to 

confirm and move to the next step, or press ON/OFF to exit.   

5.

 

“CAL.P1“ will be displayed and the scale will be ready to accept the next calibration point. The default calibration input weight 

will be 50% of the scale’s rated capacity. Use the PRINT/HOLD and UNIT keys adjust to the standard calibration weight you will 

use.  Note: do not use less than 10% of  the scale’s rated capacity.    Place the standard calibration weight on the platform 

and press   TARE   to confirm your input and move to the next step.    If “CAL.Er” is displayed, press ZERO/ON/OFF key to 

exit this mode and begin again. 

6.

 

“CAL.P2“ will be displayed and the scale will be ready to accept the next calibration point. This step is optional, but better 

linearity will be achieved by calibrating to a second point.    To skip this step, press the   TARE  key.  Otherwise, the default 

calibration input weight will be 100% of the scale’s rated capacity. Use the PRINT/HOLD and UNIT keys adjust to the standard 

calibration weight you will use.    Note: do not use less than 20% of  the scale’s rated capacity.    Place the standard 

calibration weight on the platform and press   TARE   to confirm your input and move to the next step.    If “CAL.Er” is 

displayed, press ZERO/ON/OFF key to exit this mode and begin again. 

7.

 

“CAL.P0“will be displayed again.    Remove all weight from the platform and press   TARE   to confirm and enter normal 

weighing mode. 

 

Summary of Contents for PS-103-180

Page 1: ...W W W M E A S U R E T E K N E T PS 103 Operation Manual Contents subject to change without notice Version 1 0 11 2011 ...

Page 2: ......

Page 3: ...forms poorly try moving to a different room or location Avoid using in heavy vibration or heavy airflow conditions Specifications Model PS 103 35 PS 103 68 PS 103 75 PS 103 180 PS 103 200 Max Capacity 77lb 35kg 150lb 68kg 165lb 75kg 397lb 180kg 440lb 200kg Readability 0 05lb 0 02kg 0 5lb 0 2kg 0 1lb 0 05kg 0 5lb 0 2kg 0 2lb 0 1kg Resolution 1 1750 1 300 1 1500 1 900 1 2000 Construction Epoxy paint...

Page 4: ...from right to left UNIT Weighing mode Chooses weight unit of measure Setup Calibration mode Increases the digit in the flashing data entry position by 1 TARE Weighing mode Tares the weight Setup or Calibration mode Confirms the input data and forwards to next step ON OFF ZERO Weighing mode 3 seconds Zeros the platform weight 3 seconds Powers off the scale Setup or Calibration mode Exits to normal ...

Page 5: ... The default calibration input weight will be 50 of the scale s rated capacity Use the PRINT HOLD and UNIT keys adjust to the standard calibration weight you will use Note do not use less than 10 of the scale s rated capacity Place the standard calibration weight on the platform and press TARE to confirm your input and move to the next step If CAL Er is displayed press ZERO ON OFF key to exit this...

Page 6: ...ight within 3 seconds Note When weight exceeds 10d and the variation is within 2d 50d the stable weight is displayed until less than 10d is detected on the platform 1d 1 division 1 unit of readability as noted in Specifications U5 1 Serial communication output format 0 Communication is disabled 1 Outputs stable weight unit and status when PRINT HOLD is pressed but does not receive data 2 Output gr...

Page 7: ...et reading minus decimal point unit CR LF H1H2H3 CR ETX Bytes used 8 Weight reading 1 Minus 1 Decimal point 2 or 5 Weight unit 3 Current status H1 H2 H3 U5 3 Continuously outputs the current displayed reading weight unit and current status data but will not receive data The output format is the same as U5 1 U5 4 Continuously outputs the current gross weight tare weight net weight data weight unit ...

Page 8: ...e character indicates positive weight The minus sign follows after the first digit Leading zero digits are suppressed Serial Communication Symbols LF Line Feed character hex 0AH CR Carriage Return character hex 0DH ETX End of Text character hex 03 SP Space hex 20H H1H2H3 Three status bytes see below for byte definitions p Polarity minus sign for negative weight and space for positive weight W1 W8 ...

Page 9: ...F H1H2H3 CR ETX Zero function is activated and returns to current scale status like pressing ZERO ON OFF If zero function cannot be activated it will return to current scale status Command T CR 54h 0dh Response LF H1H2H3 CR ETX Tare function is activated and returns to current scale status like pressing TARE If tare function cannot be activated it will return to current scale status Command U CR 5...

Page 10: ... 3 6V Replacement Parts Part Number Description MH12R97601G AC120V DC6V 500mA UL adaptor One Year Limited Warranty MeasureTek products covered in this manual are guaranteed to be free from defects in material and workmanship for a period of one year after date of purchase Misuse accidental damage overload alteration and improper installation are expressly excluded Any product which is determined t...

Reviews: