present readings as Relative reference value and display on secondary. Relative measurement has two
mode. one is REL

r

 = measuring value - Reference value, the other is 

REL%= (REL

r

 /Reference value)

x 100% ( press SELECT to select REL

r

 or REL% mode).

l

  While SET  button operating REL

r

 button change as 

u

 (moving right) button.

l

  Set up reference value for measurement .

1. In every function use RANGE select your range.

2. After press SET once, press SELECT twice the meter enter set up reference value for measurement.
At same time the 

pqt

 

u

 button are started.

3. Use 

pqt

 

u

 buttons set your reference value.

4. After set up, you can press SET to confirming.

l

  Press REL

r

 button more than 2 seconds the meter return to normal state.

8. SET

l

  Press this button to starting 

p

 RANGE 

q

 2nd VIEW 

t

 MAX/MIN 

u

 REL

r

 buttons. Use these

buttons set up your digits . At this condition the RANGE, 2nd VIEW, MAX/MIN, REL

r

 original functions

are disabled.

l

  Press this button more than 2 seconds backlight on and press this button again the backlight off.

The light can auto off after it on 30 seconds if does not press this button.

Display

Table 4. symbols of display

Order No.

Symbol

Description

1

Analog bar graph

2, 3,17

Negative sign

4

         OUT

Square wave output

5

Hi

Hi frequency or themocoupIe indicate

6

Battery power is weakening

7

Diodelaudible continuity function

8

REL

r

%

Relative measurement

9, 19

DC, AC, DC+AC

DC,AC, DC+AC voltage or current

10

PH+ PH-

+ peak hold, - peak Hold

11

A-H

Auto Hold

12

AVG

Average reading

13

Auto

Auto mode

14

APO

Auto power off sign

15

RS 232

Communication on annunciation

16

MAX/MIN/MAX-MIN

MAX Reading, MlN Reading /MAX-MIN Reading

20

mV/V/mA/A

Second display volt and current unit

21

Hz/kHz/MHz/

V

/k

V

/M

V

Frequency and Resistance (ohms) unit [second]

22

 ms 

 ms

Duty cycle unit, pulse width unit

23

nF/ 

m

F

Capcitance unit

24

mV/V/mA/A

Primary display volt and current unit

25

dBm

dBm annunciation

26

Hz/KHz/MHz/

V

/k

V

/M

V

Frequency and Resistance (ohms) unit (primary)

27,18

0

0

F

Temperature measurement indicate

28

RPM

Round/per minute

Figure 4. Display

Special functions

l

  Auto power off

The meter has auto power off function, in normal conditions, when the meter is power on, if any push
buttons are not used or rotary switch is not changed, it can auto power off in 30 minutes after power on.
Before power off five minutes, the audible five beepers that points out the meter will power off. In operating
state any push button is used or rotary switch is changed the time of auto power off will recount.

l

  Set up time for measurement

1. Press TIMER button to set time [ “0.00.00” appear at secondary display].

2. Press SELECT button the last digit of secondary display flash, at the same time, the 

pqt

 

u

 are

enable [The first digit of secondary display is hour, the second and third are minutes, tbe forth and fifth
are seconds].

1. Lise 

pqt

 

u

 buttons to select digit of time .

2. Press TIMER button again to confirming Complete this procedure measuring lime is starting .

l

  Set limited measurement

The meter has set up limited or down limited measurement functions, up limited and down limited setting
as following.

1. Set Up limited : power on the meter 

g

 select range 

g

 press SET 

g

 press SELECT select Up limited

seting state (secondary display “ 

 ”) press moving buttons 

pqt

 

u

 to set digits 

g

 press SET to

confirm.

2. Set Down limited : power on the meter 

g

 select range 

g

 press SET 

g

 press SELECT select Down

limited seting state (secondary display “ 

 ”) press moving buttons 

pqt

 

u

 to set digits 

g

 press

SET to confirm.

Summary of Contents for 81K

Page 1: ...s as below l Use the meter only as specified in this manual Other wise the protection provided by the meter maybe impaired l Never measure voltage while the test leads are inserted into the current in...

Page 2: ...MIN AVG MAX MIN and Present Reading on secondary display l Press this push button more than 2 seconds meter retum to normal mode l At MAX MIN state the recording time is 36 hours l While SET button op...

Page 3: ...A Second display volt and current unit 21 Hz kHz MHz V kV MV Frequency and Resistance ohms unit second 22 ms ms Duty cycle unit pulse width unit 23 nF mF Capcitance unit 24 mV V mA A Primary display...

Page 4: ...surement range is important l If the range is too high the meter will not display the must accurate measurement If the range is too low the meter will show OL on the display 2 Auto range and manual ra...

Page 5: ...ce the probes in parallel with a circuit or component when the leads are plugged into the current terminals Never test voltage when test lead plug in mA or 10A terminal Warning to wrong operation when...

Page 6: ...r on the display screen If checking point resistance fall below 50 V the beeper will sound l Hi resistance mode At normal mode press SELECT until Hi sign appear on display screen Using this function c...

Page 7: ...al speed measurement l Square Wave Output This meter can output square wave that frequency is selectable by 2nd VIEW 0 5000Hz 1 0000Hz 2 0000Hz 10 000Hz 50 000Hz 60 240Hz 74 63Hz 100 00Hz 151 50Hz 200...

Page 8: ...80 00dBm 1 0 rdg 0 01dBm Table 18 Resistance V Range Resolution Accuracy Note 800V 0 01V 8KV 0 1V Overload protection 80KV 1V 0 3 rdg 10 250V RMS 800kV 10V 8MV 100V 80MV 1kV 2 5 rdg 8 60MV 80MV 3 5 r...

Page 9: ...ads from terminals 3 Loosen the screw on battery case pull up and moce the case 4 Replace the defective battery 5 Reverse the procedure of opening to close the battery case Fuse Replacement Waring To...

Reviews: