interference. Some random digits on display but have not effect
on result of measurement.
3. At millivoltage measurement mode in order to obtain DC+AC
function, the input terminal of ADC do not employed coupling
capacitor. There for please never apply more than the double dc
or ac voltage of the rated value of this range.

l

  AC/DC millivoltage measurement

The measurement of millivoltage has three modes : DCmV / [ACmV+Hz] / dBm.
1.

Set the rotary switch to “   mV ” position.

2.

Press SELECT to select measurement mode.

3.

Connect the black test lead to “COM” terminal and the red test lead to “V 

V

 Hz” terminal.

4.

According to your need, can press REL

r

 MAX / MIN and 2nd VIEW buttons obtain relevant

measurement.

5.

Touch the probes to the test points and reading display.

6.

Primary and secondary display as Table 7.

Table 7. primary and secondary display

Press

Secondary

SELECT

Primary

[ press 2nd VIEW ]

DCmV

DCmV

[ACmV + Hz ]

ACmV

Hz/

%/

%/

ms/

ms

dBm

dBm

Hz/ACmV/DCmV/[DCmV+ACmV]

Note :
1. When enter to dBm measurement, the impedance is 600 

V

,

if you want change impedance, press RANGE to select impedance.
The impedance selectable is :
4/8/16/32/50/75/93/110/125/135/150/200/250/300/500/600/800/
900/1000/1200 of Ohms.
2. At millivoltage mode the input impedance more than 1000

V

, therefore at test leads opening state input easy caused

Figure 7. mV measurement

Warning For Current measurement

To avoid damage to meter or injury, if the fuse blows. Never attempt an in-circuit current measurement.
Where the open-circuit potential to earth is greater than 1000V. To avoid damage to the meter, check the
meter’s fuses before proceeding. Use the proper terminals, function and range for your messurement.
Never place the probes in parallel with a circuit or component when the leads are plugged into the current
terminals.Never test voltage when test lead plug in “mA” or “10A” terminal !

Warning to wrong operation when probes are plugged in to the “mA” or “10A” terminal and the rotary
switch is not at “   mA ” or “   A ” position, meter will beeper warning to wrong operation until rotary
switch at right position or probes pull out “mA” or “10A” terminals.

l

  AC/DC milliampere current measurement

The measurement of milliampere has four modes : DCmA / ACmA /
[DC mA + ACmA] / [ACmA + Hz].

1.

Set the rotary switch to “   mA ” position.

2.

Press SELECT to select measurement mode.

3.

Connect the black test lead to “COM” terminal and the red test

lead to “mA” terminal.

4.

According to your need, can press REL

r

 MAX / MIN buttons

obtain relevant measurement. Turn off power to the circuit, discharge
all high-voltage capacitors.

5.

Break the circuit path to be tested. Touch the black test lead

to more negative side of the break; Touch the red test lead to the
more positive side of the break.

Figure 8. mA Current

measurement

6.

Turn on power to the circuit, then read the display.

7.

Primary and secondary display as Table 8.

8.

Turn off power to the circuit and discharge all high-voltage

capacitors. Remove the meter and restore the circuit to normal
operation. Pull out the red test lead from “mA” terminal.
Table 8. primary and secondary display

Press

Secondary

SELECT

Primary

[ press 2nd VIEW ]

DCmA

DCmA

ACmA

ACmA

DCmA+ACmA

DCmA+ACmA

ACmA

ACmA+Hz

ACmA

Hz

l

  AC/DC Ampere current measurement

The measurement of ampere current has four modes : DCA /
ACA / [DCA+ACA] / ACA+Hz
1.

Set the rotary switch to “   A ” position.

2.

Press SELECT to select measurement mode.

3.

Connect the black test lead to “COM” terminal and the red

test lead to “10A” terminal.

4.

According to your need, can press REL

r

 MAX / MIN

buttons obtain relevant measurement.

Figure 9. Ampere current

measurement

Summary of Contents for 81K

Page 1: ...s as below l Use the meter only as specified in this manual Other wise the protection provided by the meter maybe impaired l Never measure voltage while the test leads are inserted into the current in...

Page 2: ...MIN AVG MAX MIN and Present Reading on secondary display l Press this push button more than 2 seconds meter retum to normal mode l At MAX MIN state the recording time is 36 hours l While SET button op...

Page 3: ...A Second display volt and current unit 21 Hz kHz MHz V kV MV Frequency and Resistance ohms unit second 22 ms ms Duty cycle unit pulse width unit 23 nF mF Capcitance unit 24 mV V mA A Primary display...

Page 4: ...surement range is important l If the range is too high the meter will not display the must accurate measurement If the range is too low the meter will show OL on the display 2 Auto range and manual ra...

Page 5: ...ce the probes in parallel with a circuit or component when the leads are plugged into the current terminals Never test voltage when test lead plug in mA or 10A terminal Warning to wrong operation when...

Page 6: ...r on the display screen If checking point resistance fall below 50 V the beeper will sound l Hi resistance mode At normal mode press SELECT until Hi sign appear on display screen Using this function c...

Page 7: ...al speed measurement l Square Wave Output This meter can output square wave that frequency is selectable by 2nd VIEW 0 5000Hz 1 0000Hz 2 0000Hz 10 000Hz 50 000Hz 60 240Hz 74 63Hz 100 00Hz 151 50Hz 200...

Page 8: ...80 00dBm 1 0 rdg 0 01dBm Table 18 Resistance V Range Resolution Accuracy Note 800V 0 01V 8KV 0 1V Overload protection 80KV 1V 0 3 rdg 10 250V RMS 800kV 10V 8MV 100V 80MV 1kV 2 5 rdg 8 60MV 80MV 3 5 r...

Page 9: ...ads from terminals 3 Loosen the screw on battery case pull up and moce the case 4 Replace the defective battery 5 Reverse the procedure of opening to close the battery case Fuse Replacement Waring To...

Reviews: