  Temperature measurement

1.

Set the rotary switch to “TEMP” position.

2.

Press SELECT to select Hi or normal test mode.

3.

In Hi mode using type K thermocouple to measure temperature.

4.

Plug the red leg ( + ) to “V 

V

 Hz” terminal and the black leg ( - ) to “COM” terminal and reading

display.

5.

Press SELECT to “Hi” off the reading is room temperature.

6.

Measurement range : -50

0

C -1300

0

C, -58

0

F-2502

0

F.

7.

Display : primary 

0

C, secondary 

0

F.

Figure 13.

temp. measurement

l

  Diode and continuity check

 Caution

To avoid possible damage to the meter or to the equipment under test, disconnect circuit power
and discharge all high-voltage capacitors before checking diodes.

1.

Set the rotary switch to “ 

 ” position .

2.

Connect the black test lead to “ COM” terminal and the red test lead to “V 

V

 Hz” terminal.

3.

For diode checking, touch the red test lead to the positive side of the diode and the black test lead

to the negative side. The meter can display diode voltage drop.
A good product a forward bias reading of 0.5V to 0.8V.

4.

Reverse the probes and measure the voltage across the

diode again if the diode good “OL” is displayed.

Note :

(1) Near 0V drop is displayed in both directions the diode
shorted .

(2) “OL” is displayed in both directions the diode opened.

5.

For continuity checking, while testing continuity, the

beeper will sound if the resistance falls below 50 

V

.

Figure 12. freq. and rotational

speed measurement

l

  Square Wave Output

This meter can output square wave that frequency is selectable by 2nd VIEW 0.5000Hz / 1.0000Hz /
2.0000Hz / 10.000Hz / 50.000Hz / 60.240Hz / 74.63Hz / 100.00Hz / 151.50Hz / 200.00Hz / 303.00Hz /
606.10Hz / 1.2500kHz / 1.6660kHz  / 2.5000kHz / 5.0000kHz.

1.

Set the rotary switch to “ 

 OUT ” position.

2.

Connect the black test lead to “COM” terminal and the red test lead to “V 

V

 Hz” terminal.

3.

Press SELECT to select the duty of 1% to 99%. The square wave output from “COM” and “V 

V

 Hz”

terminal or test leads.

4.

Primary and secondary display as Table 10.

Table 10. primary and secondary display

Function

Primary

Secondary

[press 2nd VIEW]

[press  SELECT]

 OUT

Hz

 %

5.

Press 2nd VIEW more than 2 seconds the meter return to

606.10Hz / 50% duty output state .

Figure 15.

square wave output

Electrical Specifications

Accuracy is specified for one year after calibration at operating temperatures of 18

0

C to 28

0

C, with

relative humidity at 0% to 75%

Accuracy specifications take the form of : ±(a% of r number of least significant digits).

Table 11. DCV

Range

Resolution

Accuracy

Note

80mV

1

m

V

800mV

10

m

V

8mV

0.1mV

± (0.05% rdg + 10)

Input impedance : 80mV - 800mV>1000M

V

80V

1mV

8V - 1000V : 10M

V

800V

10mV

1000V

0.1V

± (0.08% rdg + 10)

Table 12. ACV ( True RMS )

Range

Resolution

Accuracy

50Hz / 60Hz

50Hz - 1kHz

1kHz - 10kHz

10kHz - 20kHz

80mV

1

m

V

±(0.8% rdg + 10)

±(1.0% rdg + 10)

±(3.0% rdg + 10)

±(8.0% rdg + 10)

800mV

 10

m

V

±(0.8% rdg + 10)

±(1.0% rdg + 10)

±(3.0% rdg + 10)

±(5.0% rdg + 10)

8V

0.1mv

±(0.8% rdg + 10)

±(1.0% rdg + 10)

±(3.0% rdg + 10)

±(5.0% rdg + 10)

80V

1mV

750V

10mV

50Hz - 400Hz : ±(1.0% rdg + 5)

Input impedance : 8mV - 800mV>1000M

V

, 8V - 1000V, 10M

V

, Parallel capacitance<100pF

Summary of Contents for 81K

Page 1: ...s as below l Use the meter only as specified in this manual Other wise the protection provided by the meter maybe impaired l Never measure voltage while the test leads are inserted into the current in...

Page 2: ...MIN AVG MAX MIN and Present Reading on secondary display l Press this push button more than 2 seconds meter retum to normal mode l At MAX MIN state the recording time is 36 hours l While SET button op...

Page 3: ...A Second display volt and current unit 21 Hz kHz MHz V kV MV Frequency and Resistance ohms unit second 22 ms ms Duty cycle unit pulse width unit 23 nF mF Capcitance unit 24 mV V mA A Primary display...

Page 4: ...surement range is important l If the range is too high the meter will not display the must accurate measurement If the range is too low the meter will show OL on the display 2 Auto range and manual ra...

Page 5: ...ce the probes in parallel with a circuit or component when the leads are plugged into the current terminals Never test voltage when test lead plug in mA or 10A terminal Warning to wrong operation when...

Page 6: ...r on the display screen If checking point resistance fall below 50 V the beeper will sound l Hi resistance mode At normal mode press SELECT until Hi sign appear on display screen Using this function c...

Page 7: ...al speed measurement l Square Wave Output This meter can output square wave that frequency is selectable by 2nd VIEW 0 5000Hz 1 0000Hz 2 0000Hz 10 000Hz 50 000Hz 60 240Hz 74 63Hz 100 00Hz 151 50Hz 200...

Page 8: ...80 00dBm 1 0 rdg 0 01dBm Table 18 Resistance V Range Resolution Accuracy Note 800V 0 01V 8KV 0 1V Overload protection 80KV 1V 0 3 rdg 10 250V RMS 800kV 10V 8MV 100V 80MV 1kV 2 5 rdg 8 60MV 80MV 3 5 r...

Page 9: ...ads from terminals 3 Loosen the screw on battery case pull up and moce the case 4 Replace the defective battery 5 Reverse the procedure of opening to close the battery case Fuse Replacement Waring To...

Reviews: