background image

MPCIB 396 P6 

                                                                                                                     

 mect s.r.l.

 

    ME3025_13  03/21

                                                          

16

   

2.3 FRONT KEYS ENABLING 

The  keys  used  on  the  front  of  the  instrument  for  the  direct  sets  up  (reset,  total 
counting,  alarms  and  decimal  point)  can  be  disabled  from  the  programming  menu. 
Follow the next table. 

Table 5 

N seq. 

Key to 
press 

Appears on 
the display 

NOTES 

prog. 

  

PASS 

Touch the “prog. 

” key to get into the  

programming menu 

prog. 

  

0 0000 

Digit the personal Password. Press “prog. 

”  

to confirm. (see “Password function”) 

 

OUtPUt 

 

 

InPUt 

 

 

C.PASS 

 

 

AbtASt 

KEYS ENABLING 

prog. 

  

AbtrE 

"Reset” KEY ENABLING 

prog. 

  

On 

On = enabled, OFF= disabled 
Press  “

”  key  till  when  the  desired  function 

appears on the display  
**(press “prog. 

” to confirm) 

 

AbtrE 

 

10 

 

AbtPd 

"d.p.” KEY ENABLING (decimal point) 

11 

prog. 

  

On 

On = enabled, OFF= disabled 
Press  “

”  key  till  when  the  desired  function 

appears on the display 
**(press “prog. 

” to confirm) 

12 

 

AbtPd 

 

13 

 

Abtrt 

"

” KEY ENABLING  

14 

prog. 

  

On 

On = enabled, OFF= disabled 
Press  “

”  key  till  when  the  desired  function 

appears on the display  
**(press “prog. 

” to confirm) 

15 

 

Abtrt 

 

16 

 

AbtPr 

"Print” KEY ENABLING  

17 

prog. 

  

On 

On = enabled, OFF= disabled 
Press  “

”  key  till  when  the  desired  function 

appears on the display  
**(press “prog. 

” to confirm) 

18 

 

AbtPr 

 

19 

 

AbtSP 

 “AL” KEY ENABLING (alarms) 

Summary of Contents for MPCIB396 P6 Series

Page 1: ...ME3025_13 03 21 USER S MANUAL Series MPCIB396 P6 Bidirectional counter ...

Page 2: ......

Page 3: ...ng 18 2 6 PrESEt FUNCTION preset 19 2 7 DEFAULT PARAMETERS dEF 19 2 8 TOTAL COUNTING FUNCTION 19 2 9 MONODIRECTIONAL PULSE COUNTER FUNCTIONING 20 2 10 A B AND A B FUNCTIONING 20 2 11 MARKER PULSE FUNCTIONING 20 3 0 ALARMS option 20 3 1 ALARM SETTING 22 4 0 ANALOGUE OUTPUTS option 26 4 1 TECHNICAL CHARACTERISTICS 26 4 2 INSTALLATION OF ANALOGUE OUTPUT 26 4 3 NOTES ON THE SET UP OF ANALOGUE OUTPUTS ...

Page 4: ...MPCIB 396 P6 mect s r l ME3025_13 03 21 4 5 8 BASIC PROGRAM 39 6 0 PASSWORD FUNCTION 40 7 0 SET UP 41 8 0 NOTES 41 ...

Page 5: ...o 65535 programmable pre set offset possibility to read on 1 2 or 4 edges of frequency input serial outputs if requested analogue outputs if requested one count input and one Up Down control input two indipented count inputs A and B with A B or A B function possibility to able the marker pulse by hold terminal 1 1 TECHNICAL FEATURES Table 1 Inputs Bi directional npn pnp encoder 3 wire npn pnp ampl...

Page 6: ...an be disabled in the menu Clears count can be disabled in the menu fast exit in menu it sets up decimal point it can be disabled in the menu Used for set up readout print Used for set up Total counting function it can be disabled in the menu access at the programming functions and confirm of the selected function KEYBOARD DESCRIPTION Led AL1 alarm 1 status indication Led AL2 alarm 2 status indica...

Page 7: ...encoder power supply Terminal 18 stop counting marker pulse abilitation print if serial output requested reset total counting Terminal 19 reset counting print if serial output requested reset total counting Terminal 20 encoder marker pulse Terminal 21 encoder input 2 direction input counting if monodirectional pulse counter Terminal 23 encoder input 1 input counting if monodirectional input counti...

Page 8: ... dgnd reset input 2 hold tacca zero rs Val input 1 1 5 NPN or PNP INPUTS CONFIGURATION The hold and reset inputs can be used with npn or pnp polarity npn input short circuit JP7 and program the menu item nInP nPn pnp input short circuit JP14 and program the menu item nInP PnP The instruments are delivered with NPN inputs ...

Page 9: ...and the as indicated in diagram menu If the set up needs a number to write use the key to increase the digit which blinks and key to move the blinking digit and confirm with key If the set up needs the selection of an item use key and confirm with key Press key to go to the upper level To exit the menu press the modified parameters will be stored ...

Page 10: ...MPCIB 396 P6 mect s r l ME3025_13 03 21 10 1 7 BASE INSTRUMENT MENU DIAGRAM Notes The symbol means The symbol means ...

Page 11: ...MPCIB 396 P6 mect s r l ME3025_13 03 21 11 1 8 OPTIONS MENU DIAGRAM Notes The symbol means The symbol means ...

Page 12: ...nt is placed in the right position confirm with key 2 Get into the menu with key Program the functions of the following table to select the requested type of functioning Table 3 n seq Key to press Appears on the display NOTES 1 prog PASS Touch the prog key to get into the programming menu 2 prog 0 000 Digit the personal Password Press prog to confirm see Password function 3 OUtPUt 4 InPUt 5 CPASS ...

Page 13: ... Set analogue outputs if requested see paragraph 11 Set if desired the programming menu access code password function 12 For default parameters see default parameters paragraph 13 The unit is now ready to be used Table 4 n seq Key to press Appears on the display NOTES 1 prog PASS Touch the prog key to get into the programming menu 2 prog 0 0000 Digit the personal Password Press prog to confirm see...

Page 14: ...ARTIAL COUNTING STORED AT THE SWITCHING OFF 18 On On it memorizes the counting OFF it doesn t memorize the counting Press key till when the desired function appears on the display press prog to confirm 19 StorE 20 nInP INPUT TERMINALS SELECTION 21 prog nPn nPn reset and hold inputs have NPN polarity PnP reset and hold inputs have PNP polarity Press key till when the desired function appears on the...

Page 15: ...received at the input visualizing them as you desire The two menu items that you have to programme mean nUnEr display readout ______________ CoEnC Input pulses dEnon For a reading without correction factor is sufficient to set up nUnEr dEnon instead to add corrective constant is necessary to set up nUnEr and dEnon to get the desired value As described in CoEnC function paragraph it is possible to ...

Page 16: ... KEYS ENABLING 7 prog AbtrE Reset KEY ENABLING 8 prog On On enabled OFF disabled Press key till when the desired function appears on the display press prog to confirm 9 AbtrE 10 AbtPd d p KEY ENABLING decimal point 11 prog On On enabled OFF disabled Press key till when the desired function appears on the display press prog to confirm 12 AbtPd 13 Abtrt KEY ENABLING 14 prog On On enabled OFF disable...

Page 17: ...provided Tabella 6 N seq Key to press Appears on the display NOTES 1 prog PASS Touch the prog key to get into the programming menu 2 prog 0 0000 Digit the personal Password Press prog to confirm see Password function 3 OUtPUt 4 InPUt 5 C PASS 6 AbtASt 7 CnFnor TERMINALS CONFIGURATION 8 prog n 19 TERMINAL 19 CONFIGURATION 9 prog rESEt Terminal 19 rESEt reset counting Print print if serial output re...

Page 18: ...much as possible the encoder resources Infact the bi directional encoder produces two waves dephased of 90 The reading of an edge every four allows to visualize the encoder revolution pulses this function is obtained with the set up of CoEnC F1 fig C By the CoEn set up it is possible to read two or four edges getting double or quadruple readings about the encoder revolution pulses To double the re...

Page 19: ...ion look the following table BE CAREFUL all previous programmed values will be lost Table 7 n seq touch key Appears on the display NOTES 1 prog PASS Press prog key to get into the programming menu 2 prog 0 0000 Digit the personal password confirm with prog 3 OUtPUt 4 InPUt 5 C PASS 6 AbtASt 7 CnFnor Terminal configuration if serial output are requested 8 dEF DEFAULT PARAMETERS 9 prog On Touch the ...

Page 20: ... input in subtract counting mode The two used inputs are input 1 at terminal 23 for up counting and input 2 at terminal 21 for down counting 2 11 MARKER PULSE FUNCTIONING The fifth encoder wire called marker pulse is abled to zero the visualization just when the hold terminal is programmed as marker pulse abilitation When terminal 18 is active the instrument is abled to zero the visualization when...

Page 21: ...tivated or deactivated Besides for SP1 and SP2 it is possible to programme delay time or hysteresys see figure D After reset the display shows the preset value Hy Hy Hy Hy Sp2 Sp1 figure D THRESHOLD SOGLIA The output changes when the counting crosses the SP1 set point The output can be normally activated or deactivated Besides for SP1 it is possible to set up delay time or hysteresys see figure E ...

Page 22: ...r a programmed time shows 1100 on the display and restarts until reaches 0 again Example 2 SP1 1000 PrESEt 100 Counting decrease till 0 switch the relay for a programmed time shows 900 on the display and restarts until reaches 0 again OFFSET THRESHOLD This function is available only for alarms 2 and 3 The offset value is respect setpoint 1 Example Alarm 1 Threshold Sp1 1000 Alarm 2 dELtA Sp1 20 98...

Page 23: ...he desired item by key and confirm with prog 8 tIPOAL 9 rELE AL1 CONTACT CONFIGURATION 10 prog nA n A relay normally open n C relay normally closed Select the desired item by key and confirm with prog 11 rELE 12 SP 1 First trigger set point SET UP 13 prog 0 00000 Set up the SP1 value Confirm by prog 14 SP 1 15 SP 2 Second trigger set point SET UP Program only if the functioning of the windowed ala...

Page 24: ...he desired item by key and confirm with prog 28 tIPOAL 29 rELE ALARM 2 FUNCTIONING CONFIGURATION 30 prog NA n A relay normally open n C relay normally closed Select the desired item by key and confirm with prog 31 rELE 32 SP 1 First trigger set point SET UP 33 prog 0 00000 Set up the SP1 value Confirm by prog 34 SP 1 35 SP 2 Second trigger set point SET UP Program only if the functioning of the wi...

Page 25: ...e menu see para SETTING to change the set value After to have configured the alarms it is possible to get in the change of set point by the front key AL See the next table Table 9 N SEQ Press key appears on display NOTES 1 AL AL 01 Press AL key to get into the set up alarms set point 2 prog SP 1 First set point alarm 1 3 prog 0 00000 Write the alarm 1 relative number confirm with prog 4 SP 1 5 SP ...

Page 26: ... analogue output the user will have to set up the two reading values IS and FS and the corresponding output values ISO and FSO It is necessary to take account of the fact that for reading values smaller or greater than those programmed the analogue output will not rise above the values set by the programming parameters ISO and FSO output values For greater clarity please check with the paragraph N...

Page 27: ...f the analogue output ISO The programming parameter ISO is programmed depending on the type of output chosen Consequently we can obtain ISO 00 000 V if voltage output ISO 00 000 mA if ammeter output for output 4 20 mA no programming is necessary FSO end of the output scale is the value of the analogue output which coincides with the programmed number at the item FS Digit at the FS item the display...

Page 28: ...ed item with key and confirm with prog key 8 SEL OA 9 IS BEGINNING OF READING SCALE 10 prog 0 00000 write the reading value which coincides with ISO confirm with prog 11 IS 12 FS END OF READING SCALE 13 prog 1 00000 write the reading value which coincides with FSO confirm with prog 14 FS 15 ISO BEGINNING OF ANALOGUE OUTPUT SCALE 16 prog 000 00 write the output value which coincides with the readin...

Page 29: ...rument with the following calibration READING 100 AMMETER OUTPUT 4 mA READING 10000 AMMETER OUTPUT 20 mA The instrument parameters should be programmed as follows PARAMETERS OUt AN SEL OA C 4 20 IS 100 FS 10000 ISO not necessary to program FSO not necessary to program with display below 100 the analogue output is fixed to 4 mA with display above 10000 the analogue output is fixed to 20mA Program i...

Page 30: ...nked with each other 31 max by this line transmission All messages are sent and received by means of an ASCII protocol Table 12 SERIAL CHARACTERISTICS baud rate From 1200 to 38400 programmable start bit 1 bit Length 8 bit Stop 1 bit Parity no To use the instruments MPCIB396 P6 models with RS485 or RS422 serial output you must follow figure G for RS232 serial output follow figure F program the inst...

Page 31: ...te 2400 2400 baud rate 4800 4800 baud rate 9600 9600 baud rate 19200 19200 baud rate 38400 38400 baud rate Press the key key until appears the requested baud rate confirm with prog 9 bAUd 10 Addr INSTRUMENT ADDRESS 11 prog 001 Input the instrument address with a number between 001 and 099 confirm with prog 12 Addr 13 Exit Reset measure Procedure to exit to programming environment to modify the num...

Page 32: ...MPCIB 396 P6 mect s r l ME3025_13 03 21 32 25 14 13 1 Out DAT I In DAT I GND Host computer L L L L GND L L GND Figure F RS232 Figure G RS485 ...

Page 33: ...when the data reception is complete the instrument depending on the information received can behave in the following ways 1 If the data string received presents errors which do not allow address identification then the instrument cannot reply and rejects the information received 2 The string has a correct address code but it detects other errors in this case the instrument transmits the ASCII code...

Page 34: ... to host in response to preceding example STX P R blank blank blank blank 0 1 0 0 ETX BCC 02 50 52 20 20 20 20 30 31 30 30 03 00 cod ascii The MPCIB396 P6 after having transmitted the string with the data requested from the host computer awaits the reply confirming the result of the transmission 1 The host computer replies in ASCII NACK retransmit the message The MPCIB396 P6 retransmits the data s...

Page 35: ...ormation received can behave in four different ways 1 If the data string received presents errors which do not allow address identification then the instrument cannot reply and rejects the information received 2 The string has a correct address code but detects other errors in this case the instrument transmits the ASCII code NACK not understood and rejects the information received 3 The received ...

Page 36: ...ead write ASCII 999999 B1 B2 B3 SP2 AL1 AL2 AL3 Read write ASCII 999999 H1 H2 H3 HY AL1 AL2 AL3 Read write ASCII 0 250 D1 D2 D3 Delay AL1 AL2 AL3 Read write ASCII 0 25 0 W1 W2 W3 Alarm status word Read write Hexadecimal see paragraph The code is composed by the letter followed by the number of the alarm to program 5 5 TRANSMISSION OF HEXADECIMAL VALUES Some values must be transmitted in a 4 digit ...

Page 37: ...the instrument in hold state D1 gets the H value while D2 keeps staying blank 5 6 ALARM SETTING W status word MPCIB396 P6 series instruments can have 3 alarms The relevant codes to program are A reading writing of SP1 B reading writing of SP2 H reading writing of hysteresis D reading writing of delay time W reading writing of the status word of the alarm The code is followed by a number indicating...

Page 38: ..._13 03 21 38 5 7 AR STATUS WORD The AR status word allows to configure the front keys enabling by serial line the terminals configuration the kind of encoder counting You can see the status word description in the following picture ...

Page 39: ...e set point of an instrument by an host computer Set up baud rate 9600 and address 01 for the functioning of the program on error goto 20 cls open com1 9600 n 8 1 for random as 1 print 1 chr 4 0 0 1 1 R O chr 5 print waiting for answer cls a input 13 1 b mid a 5 7 print print read b end 20 print no answer resume ...

Page 40: ...n or number that is flashing 2 false Password number The user can only see the programmed numbers but cannot modify them Table 15 n seq Touch key Appears on the display REMARKS 1 prog PASS Touch the prog key to get into the programming menu 2 prog 0 00000 confirm with prog 3 OUtPUt 4 InPUt 5 C PASS PERSONAL PASSWORD 6 prog 0 00000 Input a Password number between 0 and 9999 confirm to prog 7 C PASS...

Page 41: ... digit forward right 5 prog SP 1 The value is stored and the display moves back to the selected item 8 0 NOTES The instrument does not have a power on switch and an internal fuse but it immediately switch on when the correct voltage is applied see the operating voltage on the instrument label Keep the power line separate from the signal lines For security reasons it is necessary to provide externa...

Reviews: