Chapter 4 Operation
- 21 -
When the bladder projections are like what is shown in Fig. 4.19a, b, c and d, please adjust the
position and angle of the Main Unit and re-measure.
Fig. 4.19a Incorrect Position
Fig. 4.19a: The bladder projection is beyond the center of the testing area. It is under the
center of the testing area, the measurement result is inaccurate. Adjust the angle and
position of the Main Unit to the lower side of the patent and re-measure it.
Fig. 4.19b Incorrect Position
Fig. 4.19b: The bladder projection is beyond the center of the testing area. It is above the
center of the testing area, the measurement result is inaccurate. Adjust the angle and
position of the Main Unit to the upper side of the patent and re-measure it.
Fig. 4.19c: Incorrect Position
Fig. 4.19c: The bladder projection is beyond the center of the testing area. It is on the left of