background image

11

10

If There Is No Vacuum 

If you do not feel any vacuum at the breast or if the pumping

action is too weak:

Check that the breastshield forms a complete seal
around the breast.

Check to see that the membrane is securely attached to 
the valve on the underside of the breastshield. Ensure that
membrane is clean and that the flap is moveable and not 
dried onto the valve head.

For single pumping, check whether the second connection
on the pump has been closed. 

Check that the gaskets are clean and dry.

Check with fresh batteries or electrical if not powering unit 
with batteries.

Check that the connector tube is firmly plugged into the 
breastshield.

* See page 4, “#3 Air-Dry” for further information on drying tubing.

Troubleshooting

If milk accidentally gets into the tubing/motor, immediately 
follow these steps:

1

1

Remove tubing from the motor. Wipe
milk off the motor with a paper towel
or tissue. Run motor for at least 30
seconds to clear out any milk that
remains inside.

2

2

Take special care to clean the 
vacuum ports in the front sides of 
the motor.

Start by removing the covers; turning
in the direction of the arrow (counter
clockwise) and pull out.

3

3

Remove the gasket(s). Wash 
gasket(s), tubing* and cover(s) with
warm soapy water. Rinse in clear
water and dry thoroughly.

NOTE:

Only the side with potential 

overflow needs to be cleaned.

4

4

Wipe the area on the motor unit
under the gasket and on the bottom
around the vent hole.

Replace the dried gasket(s),
tubing* and cover(s).

Put the tubing back on the breast
shield and resume pumping. When 
pumping is completed, repeat 
these steps.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for Double Select

Page 1: ...medela com www medela com Instructions Assembly Instructivo de uso y armado Printed in the USA Medela Lactina and Symphony are registered trademark of Medela Inc Double Select Pump Save Quick Clean a...

Page 2: ...a damaged cord or plug if it is not working properly if it has been dropped or damaged or dropped into water 5 Keep the cord away from heated surfaces 6 Do not use outdoors 7 Plug the transformer into...

Page 3: ...her For Quicker Cleaning Medela offers Quick Clean Micro Steam Bags for daily kit disinfection Uses Collect breastmilk during short term absences from the baby Maintain a milk supply if breastfeeding...

Page 4: ...t Freshly 4 hours 24 hours 5 7 days 3 4 6 12 expressed at 66 72 F at 59 F at 32 39 F months 3 months breastmilk 19 22 C 1 15 C 1 0 4 C 2 at 0 F 19 C 3 Thawed Do not Do not 24 hours 4 Never Never breas...

Page 5: ...teries with fresh ones as soon as possible NOTE The batteries do not need to be removed to use electricity to power the pump Electric Operation 1 1 Plug the metal fitting on the end of the transformer...

Page 6: ...ping it may take about 10 15 minutes to collect milk from both breasts Single pumping may take 10 15 minutes per side The following steps of caution should be followed to prevent overflowing milk into...

Page 7: ...ation on drying tubing Troubleshooting If milk accidentally gets into the tubing motor immediately follow these steps 1 1 Remove tubing from the motor Wipe milk off the motor with a paper towel or tis...

Page 8: ...rchase In the event of a defect Medela will repair or at Medela s option replace this product without charge for such replacement parts or labor Purchaser shall bear all expense for returning this pro...

Page 9: ...sicas de seguridad como las siguientes LEA TODAS LAS INSTRUCCIONES ANTES DE USAR LA BOMBA DE LACTANCIA PELIGRO Para reducir el riesgo de electrocuci n 1 Siempre desconecte los dispositivos el ctricos...

Page 10: ...pieza m s r pida Medela ofrece las bolsas Quick Clean Micro Steam para la desinfecci n diaria Aplicaciones Colectar leche materna durante los per odos breves que est lejos del beb Mantener un suminist...

Page 11: ...ara preservar su frescura Almacenamiento de la leche materna Consulte con el hospital para conocer las instrucciones de almacenamiento espec ficas Al congelar la leche no llene los recipientes a m s d...

Page 12: ...raer las bater as para utilizar la bomba con electricidad Operaci n el ctrica 1 1 Inserte la clavija met lica del extremo del cord n del transfor mador en el orificio localizado en la parte trasera de...

Page 13: ...e puede tardar de 10 a 15 minutos para extraer la leche de ambos senos El bombeo sencillo puede tardar de 10 a 15 minutos por lado Observe estas precauciones para evitar que fluya leche al motor Tenga...

Page 14: ...Identificaci n y resoluci n de problemas Si la leche entra accidentalmente en la manguera o el motor siga de inmediato estos pasos 1 1 Separe la manguera del motor Limpie la leche del motor con una t...

Page 15: ...de la fecha de compra En caso de que el producto tuviera alg n defecto Medela reparar o reemplazar el producto a discreci n propia sin cargo alg n por piezas de repuesto o mano de obra El comprador d...

Page 16: ...DoubleEase Apart eps 28 PARTS PIEZAS 1 2 3 4 5 7 8 9 10 11 6...

Reviews: