Troubleshooting
XI
If the motor is
not running
Check
• mains power is present
• the device is switched on
If there is low
or no suction
Make sure
• the breastshield forms a complete seal around the breast.
• all connections of the kit and breastpump are secure.
• the white valve membrane is laying flat against the yellow valve
head. Check that the valve head and membrane are clean and
not damaged. See «Assembly Instructions».
• the ends of the tubing are fit securely into the back of the
breastshield and into the opening on cylinder holder.
• the rubber seal is correctly attached onto piston.
• the piston/cylinder is correctly attached onto cylinder holder or
breastshield.
• the cylinder holder including piston/cylinder is assembled pro-
perly onto the Lactina.
• If single pumping, make sure the port cap is inserted into the
unused opening of cylinder holder.
Take care not to kink or pinch tubing.
If there is an
overflow
If milk accidentally gets into the tubing
• Turn off the pump and unplug from power source.
• Use a damp (not wet) cloth to wipe the pump.
• Refer to cleaning instructions «Tubing care and Breastpump
care».
• If necessary clean the pump kit.
• When components are completely dry, reassemble
breastpump and kit.