• Patient leakage current.
• The dialectric strength between live parts and the housing. This
insulation shall be double insulation
The allowable values are defined according to the chart below:
NC:
Normal Condition
SFC:
Single Fault Condition
U:
Reference voltage
7. Terms of Guarantee
Medela guarantees the perfect quality and finish of each Lactina
®
breastpump for a period of 3 years from the factory delivery date During
this period, all faulty components will be replaced free of charge This
guarantee does not extend to components which are subject to wear
and tear or which have been damaged due to incorrect use
Further conditions according to International Standard IEC 6011, Item
682b:
The manufacturer may only be held responsible for inadequate safety,
reliability and performance of the unit if
– assembly
– additional items
– adjustments
– alterations or repairs
have been carried out by persons authorized by the manufacturer and
– the electrical installation within the room concerned satisfies IEC
requirements (in the course of preparation).
– the unit is used as specified in the instructions for use
– accessory kits used on Lactina breastpumps are authorized by
Medela, Inc
15
Test
Allowable Value
Chapter IEC 601-1
Enclosure leakage current
NC
SFC
19.1
0,1 mA
0,5 mA
19.2
19.3
Patient leakage current
DC
NC
SFC
19.1
0,1 mA
0,5 mA
19.2
AC
NC
SFC
19.3
0,1 mA
0,5 mA
19.4
Dialectric Strength
50 <U ≤150V
3000V
20.1
(double insulation)
20.3
150 <U ≤250V
4000V
20.4
(double insulation)